Specify a monitor location
You can associate the monitor with a specific location. The location appears in the Device Status
area.
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Other tab.
3. In the Location ID box, touch
and enter up to 20 alphanumeric characters.
4. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Enable monitor profile changes
You can allow clinicians to change the active profile on the monitor. Available profiles are
Monitor, Spot Check, and Triage. When this option is enabled, clinicians can change the
name of the profile as well.
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Other tab.
3. Select Allow profile change.
4. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify power line frequency
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Other tab.
126 Advanced settings