Specify clinical data settings – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 139

initiates the search when the clinician enters or
scans the ID from the Clinician tab. Returned
clinician information populates the Device Status
area and fields on the Clinician tab.
Select
Require password
to require clinicians
to enter their password, in addition to ID, on the
Clinician tab. The monitor uses the ID and
password combination to query the network for
clinician information.
Clear clinician information on manual save
Specify that the monitor clears the selected
clinician after a clinician manually saves
measurements from the Home tab. Clinician
information clears from the Clinician tab and the
Device Status area.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify clinical data settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab appears.
2. Touch the Data Management tab.
3. Touch the Clinical Data tab.
4. Specify settings.
Setting
Action/Description
Automatically send on manual save
Select this option to specify that measurements are
sent to the network when a clinician saves
measurements on the Home tab.
Delete readings after successful send
Select this option to specify that measurements are
deleted from the monitor after they are successfully
sent to the network. Sent measurements do not
appear in the Review tab.
Emulate Spot Vital Signs LXi
Select this option to specify that clinical data sent
to the network appears as Spot Vital Signs LXi data
at the network.
5. Do one of the following:
•
To continue in Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Directions for use
Advanced settings 133