Braun thermoscan® pro 4000 thermometer and dock, Braun thermoscan, Pro 4000 thermometer and dock – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 91

background image

5. The monitor sounds a tone when the final temperature is reached (in approximately 10 to 13

seconds). The temperature frame continues to display the temperature in degrees Fahrenheit

and degrees Celsius even after the probe is returned to the probe well.

Note

To switch to the Direct mode, touch

after the Predictive

mode measurement is acquired. The temperature frame (in the

lower-left corner) changes to "MODE: Direct..." as it switches to the

Direct mode. The monitor sounds a tone to indicate the start of a

Direct measurement. Once you are in the Direct mode, continue to

separate the buttocks and hold the probe in place throughout the

measurement process.

Note

The monitor does not retain Direct mode temperatures in memory.

Therefore, it is important to note the temperature before removing

the probe from the measurement site and then manually record it in

the patient record.

6. Remove the probe after the temperature measurement is complete and firmly press the eject

button on the top of the probe to release the probe cover.

7. Return the probe to the probe well.
8. Wash your hands to reduce the risk of cross-contamination.

Braun ThermoScan

®

PRO 4000 thermometer and dock

The thermometer and dock enable you to transfer an ear temperature measurement to the monitor.

The dock also charges the thermometer battery.

Read the thermometer manufacturer's directions for use before attempting to configure, use,

troubleshoot, or maintain the thermometer.

WARNING Liquids can damage electronics inside the thermometer. Prevent

liquids from spilling on the thermometer. If liquids are spilled on the thermometer,

dry off the thermometer with a clean cloth. Check for proper operation and

accuracy. If liquids possibly entered the thermometer, remove the thermometer

from use until it has been properly dried, inspected, and tested by qualified service

personnel.

CAUTION Probe covers are disposable, nonsterilized, and single-use. The

thermometer is also nonsterilized. Do not autoclave the thermometer and probe

covers. Ensure that probe covers are disposed of according to facility requirements

or local regulations.

Directions for use

Patient monitoring 85