The following actions will return the monitor display to the active state:

Touch the screen

Remove the temperature probe from the probe well

Attach the SpO2 sensor to a patient

Press

Standby

The monitor is plugged into a power outlet, but the sensors and the display do not operate.

Note

Because power is still available to charge the battery and power the monitor, the

monitor is in Standby mode.

The monitor remains in Standby mode until you press

. Settings for this mode can be changed

in the Advanced Settings Display tab.

Power up the monitor

The monitor runs a brief diagnostic self-test each time it powers up.

WARNING Equipment failure risk. The monitor includes a fan that circulates air

through the device. If the fan does not run when you power up the device, remove it

from use and inform qualified service personnel immediately. Do not use the

monitor until the problem is corrected.

WARNING To ensure patient safety, listen for two audible indicators (a piezo

beeper and a speaker tone) and watch for visual alerts at power-up. Correct any

system errors before using the monitor. In addition to the audible indicators, the

monitor LED light bar illuminates to alert you of alarms. Amber indicates a low-level

alarm. Flashing amber indicates a medium-level alarm. Flashing red indicates a

high-level alarm.

WARNING Always observe the monitor during power-up. If any display fails to

illuminate properly, or if an error code displays, inform qualified service personnel

immediately, or call your nearest Welch Allyn Customer Service or Technical

Support facility. Do not use the monitor until the problem is corrected.

CAUTION Always use the monitor with an adequately charged and properly

functioning battery. For continuous monitoring, always connect to AC power.

CAUTION Use only a Class I (grounded) AC power cord for powering this monitor.

Press

to power up the monitor.

Following a successful self-test, the monitor displays the Welch Allyn logo, the LED light bar

(located on the handle) flashes, and a power-up tone sounds. The startup screen then appears

with the following banner across the bottom.

If a system error is detected, the monitor becomes inactive until you press

or until the

monitor shuts down automatically. The monitor displays a system fault message that contains

34 Startup

Welch Allyn Connex

®

Integrated Wall System