Specify advanced temperature settings – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 134

If
Display MAP
is selected, specify which
numerics are primary in the NIBP frame. On the
Home tab, clinicians can touch the NIBP frame to
toggle between views.
Default patient type
Select a default patient type for this monitor. The
patient type shows in the Patient frame on the
Home tab.
In the Patients tab on the Summary tab, clinicians
can change the displayed patient type from the
default patient type that you set here.
Tube type
Select the number of tubes that are connected to
the NIBP cuff that is used with this monitor. If you
select
1 tube
, the only algorithm available for
selection is
Step
.
Unit of measure
Select the NIBP unit of measure for display.
Allow interval program changes
Enable clinicians to modify interval program
settings from the Intervals tab.
Algorithm and Cuff inflation target (CIT)
Select the default algorithm used to determine
NIBP measurements.
If you select the
Step
algorithm, touch
and
enter a default cuff inflation target for each type of
patient. In the Patients tab on the Summary tab,
clinicians can change the CITs from the default
CITs that you set here.
5. Do one of the following:
•
To continue in the Advanced Settings, touch another tab.
•
To exit the Advanced Settings and return to the Home tab, touch Exit.
Specify advanced temperature settings
1. Access the Advanced Settings.
a. Touch the Settings tab.
b. Touch the Advanced tab.
c. Enter the Advanced settings code.
d. Touch OK.
The General tab will appear.
2. Touch the Parameters tab.
3. Touch the Temperature tab.
4. Specify settings.
Setting
Action/Description
Unit of measure
Select primary units of measure for the
temperature display on the Home tab
Display temperature conversion
Select this to display primary units of measure and
secondary units of measure for the temperature
display on the Home tab.
Default SureTemp Plus site
Select the default site for SureTemp
measurements. The default site applies when
clinicians power up the monitor and each time
128 Advanced settings
Welch Allyn Connex
®
Integrated Wall System