Interval nibp measurement – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 80

1. Properly size the blood pressure cuff and position it around the patient’s bare upper arm.
2. Touch Start to take a measurement.
Interval NIBP measurement
The monitor can take NIBP measurements automatically based on intervals you choose.
The Intervals tab provides all interval features.
From this tab, you can do the following:
•
Configure intervals
•
Turn off intervals
•
Configure the monitor to print automatic measurements as they are completed
When the measurement is complete, the NIBP frame displays the measurement until the next
measurement is complete.
Note
During intervals, each automatic and manual save of patient measurements clears
all measurements from Manual parameters frame.
The
button changes to a timer (
), which counts down to the next automatic
measurement.
Automatic measurements continue until you turn off intervals.
WARNING Patient harm risk. Do not use intervals on neonates out of earshot.
Verify that audio can be heard from where you intend to be.
Automatic intervals
You can configure the monitor to take automatic NIBP measurements at consistent intervals.
Note
An alarm does not turn off intervals. Subsequent automatic measurements continue
to occur as scheduled.
Start automatic intervals
Follow these steps to configure the monitor to take NIBP measurements at consistent intervals.
74 Patient monitoring
Welch Allyn Connex
®
Integrated Wall System