Caviwipes™ or sani-cloth® plus – Welch Allyn Connex Integrated Wall System - User Manual User Manual
Page 110
Clean the wall system (excluding handle cradles and
accessories)
WARNING Electric shock hazard. Before cleaning the wall system, disconnect
the AC power cord from the power outlet.
WARNING Electric shock hazard. DO NOT autoclave the wall system or
accessories. The wall system and the accessories are not heat-resistant.
WARNING Liquids can damage electronics inside the wall system. Prevent
liquids from spilling on or dripping into the wall system.
If liquids are spilled on or drip into the wall system:
1. Power down the wall system.
2. Disconnect the power plug.
3. Remove the wall system from the wall.
4. Remove battery pack from the wall system.
5. Dry off excess liquid from the wall system.
Note
If liquids possibly entered the wall system, remove the wall system
from use until it has been properly dried, inspected, and tested by
qualified service personnel.
6. Reinstall battery pack.
7. Mount the wall system on the wall. (See the Setup section of this directions for use.)
8. Power on the wall system and verify that the wall system functions normally before
using it.
Clean on a routine basis according to your facility's protocols and standards or local
regulations. If the monitor is on, lock the display and disconnect the AC power cord.
The following agents are compatible with the wall system:
•
CaviWipes
™
(see Caution below)
•
Sani-Cloth
®
Plus
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach solution
CAUTION Some components and accessories of the wall system
require special care during cleaning. To ensure optimal functioning
and availability of specific components and accessories, use only
the cleaning agents noted and the processes described for these
items presented later in this section.
Note
Disinfect according to your facility's protocols and standards or
local regulations.
CaviWipes™ or Sani-Cloth® Plus
1. Using CaviWipes™ or Sani-Cloth® Plus, wipe the surface of the monitor to remove all debris.
2. Allow the monitor surface to dry for a minimum of 10 minutes before using the monitor.
104 Maintenance and service
Welch Allyn Connex
®
Integrated Wall System