beautypg.com

Configure sphb alarms, Set sphb averaging mode, Pulse rate frame – Welch Allyn Connex Integrated Wall System - User Manual User Manual

Page 101

background image

Note

A range of sensors is available for different patient sizes and

measurement sites. Consult the sensor manufacturer's instructions

for selecting the correct sensor.

5. Confirm that the monitor displays SpHb or SpHbv data within 160 seconds of connection to the

patient.

WARNING Patient injury risk. Incorrect sensor application or

excessive duration of sensor use can cause tissue damage. Inspect

the sensor site periodically as directed in the sensor

manufacturer's instructions.

While SpHb is being measured, the displayed SpO2 and pulse rate are derived from the same

sensor. If SpO2 is not available, the pulse rate is derived from NIBP.

Detaching the sensor during a measurement triggers an alarm.

If SpHb is being measured continuously on a patient for an extended period, change the sensor

location at least every three hours or as indicated by the sensor manufacturer's instructions.

Configure SpHb alarms

Follow these steps to set alarm limits for SpHb measurements.

1. Verify that you are using the Monitor profile, which contains the Alarms tab.
2. Touch the Alarms tab.
3. Touch the SpHb tab.
4. Enter the desired upper and lower alarm limits for SpHb using the up/down arrow keys or the

keypad.

5. Touch the Home tab.

The new alarm settings appear in the Alarm Limit control button.

Set SpHb averaging mode

Touch

in the SpHb frame.

The SpHb frame displays the current mode.

Pulse rate frame

The pulse rate frame, located in the upper right of the Home tab, displays data, information, and

controls used in reading pulse rates.

Typically, the displayed pulse rate is derived from the SpO2 sensor. If SpO2 is not available, the

pulse rate is derived from NIBP.

WARNING Inaccurate measurement risk. Pulse rate measurements generated

through the blood pressure cuff or through SpO2 are subject to artifact and might

not be as accurate as heart rate measurements generated through ECG or through

manual palpation.

Directions for use

Patient monitoring 95