Reset (pause or turn off) audio alarms
Audio alarm characteristics
•
After you reset an audio alarm, some tones do not return, but others return after a pause
interval if the condition that caused the alarm persists. Settings in the Advanced tab determine
the length of the pause interval.
•
If a new alarm condition occurs during a pause interval, a new audio tone occurs.
•
If an audio alarm is not paused or turned off after a period of time, a buzzer accompanies the
tone.
Pause or turn off an audio alarm
1. In the Device Status area, touch
.
•
Visual indications remain in the parameter frame until the condition is corrected or until
the next measurement is taken.
•
In the Device Status area, if the icon changes to
and the message remains, the
timer counts down and the audio tone returns after a pause interval. You can touch
again to restart the timer.
If you responded to an NIBP alarm and multiple NIBP limits have been exceeded, the first
audio tone and message go away, but another NIBP limit message shows with a
countdown timer. A new NIBP audio tone sounds after the countdown unless you touch
to dismiss each remaining NIBP limit message.
2. If multiple alarms are active, a multiple alarm toggle will appear in the Device Status area.
Respond to multiple alarms as follows:
a. Touch
in the Device Status area. (See note below.)
b. Read the alarm message for the second alarm.
c. Touch
.
d. Continue to touch multiple alarm toggle buttons and to reset tones until you have read all
of the messages.
Note
The multiple alarm toggle button will display the number of
active alarms inside the alarm icon. A set of dots indicating the
display order of alarms from highest (left) to lowest (right)
priority (as well as the most recent in the case of multiple
alarms of the same priority) will appear below it.
62 Alarms