Message
Possible cause
Suggested action
persists, move the patient and
the monitor to a cooler
location.
Device shutdown is not
available at this time.
Device cannot perform an immediate
shutdown
Touch
OK
, wait, and retry.
Advanced settings unavailable Sensors are taking measurements
Stop continuous
measurements.
A physiological alarm condition is active
Respond to or reset the alarm.
Spot Check measurements have not been
saved
Save the measurements.
Unable to load language.
Chinese did not load
Power down and restart the
monitor.
Unexpected restart occurred.
Call for service.
A system error caused the monitor to restart. Call for service.
Battery power manager messages
Message
Possible cause
Suggested action
Low battery 5 minutes or less
remaining. (High-priority alarm)
Battery power is extremely low
Connect the monitor to AC
power. (If not connected to
AC power, the monitor powers
down when AC power is
depleted.)
Low battery 30 minutes or less
remaining.
Battery power is low
Touch the alarm icon to
dismiss or connect the
monitor to AC power.
Powering down. Call for
service.
Power manager or battery faults have
occurred
Call for service.
Battery is absent or faulty.
There is no battery in the monitor
Insert a battery.
The battery is faulty
Replace the battery.
Device is operating in battery
mode.
The AC power cord has been disconnected Touch
OK
to dismiss or
connect the monitor to AC
power.
Configuration Manager messages
Message
Possible cause
Suggested action
Unable to load configuration;
using factory defaults.
A configuration load error occurred
Call for service.