Pulse / frequency output

Function

When set as frequency output: Mass and volume flow
rate, flow velocity, temperature, concentration, con-
centration flow, density, diagnosis values

When set as pulse output: volume, mass, concentra-
tion (e.g. 1 pulse / m3 or / kg)

active or passive mode, dependent on the output /
input assemblies, see section 9.8

Op. values and load
rating

Active

f

a

10 kHz: I

a

20 mA / f

a

100 Hz: I

a

100 mA

Unom 24 V DC / U0 1.5 V @ 10 mA

NAMUR to EN 60947-5-6 (operating data as for “passive“)

Pulse rate

0...10 kHz, scaleable (overflow up to fmax

a

12 kHz)

Pulse width

0.05 ...2000 ms (autom., symm. or settable)

Forward / reverse
measurement

Direction identified via status output, see below

Time constant

0...100.0 s, settable as required

Low-flow cutoff

Value: 0.0...20.0 %

of Q100% settable

as required

Hysteresis: ± 0.0...19.9 %

Status output

Function

Error in device, application error, out of specification
error, polarity, overrange, counter preset

active or passive operation, dependent on output /
input assemblies, see section 9.8

Op. values and load
rating

Active

U

a

24 V DC/ I

a

100 mA / U0

a

1.5 V @ 10 mA

Passive U

a

32 V DC/ I

a

100 mA / U0

a

1.5 V @ 10 mA

NAMUR to EN 60947-5-6 (operating data as for “passive“)

Time constant

0...100.0 s, settable as required

Control input

Function

hold outputs, set outputs to zero, stop cpunters, count-
er reset, error reset, range change,

active or passive operation, dependent on output /
input assemblies, see section 9.8.

Op. values and load
rating

Active

Inom = 16 mA / Unom = 24 V DC

Passive U

a

32 V DC/ / Uon

>

19 V DC / Uoff

<

2.5 V DC

NAMUR to EN 60947-5-6 (operating data as for “passive“)

Internal electronic
counters

Number

3, settable independent of one another

Measured variable

Total mass, volume or concentration

Function

Sum + or – counter and preset counter

Time constant

0 - 100.0 s, settable as required

Low-flow cutoff

Value: 0.0 - 20.0 %

of Q100%, settable as
required

Hysteresis:± 0.0 - 19.9 %.