The above data are provided as a rough guide only.

Driver = White / Brown

Sensor A = Orange / Black. Sensor B = Grey / Blue

RTD = Red / Purple (108

at 20°C if PT100, 540

at 20°C if PT500). Compensation leg =

Purple / Yellow

Resistance values outside these values could indicate a circuit failure. Meter may be in
start-up or have measuring errors.

All circuit should be isolated from ground (meter case) and each other:

>

20M

at 20°C. If

circuits are shorting to ground, the meter may be in start-up.

Caution:

If the measuring tube/s fail, process fluid can leak into the outer casing. If this happens,
depressurise the meter and remove it from the process line as soon as it safe to do so.

The above data are provided as a rough guide only.

Driver = Black and Grey.

Sensor A = White and Yellow. Sensor B = Green and Purple.

RTD = Red and Blue (530...550 ) at ambient temperature

Measuring Tube Strain Gauge = 420...560

Resistance values outside these values could indicate a circuit failure. Meter may be in
start-up or have measuring errors.

All circuit should be isolated from ground (meter case) and each other: >20M . If circuits
are shorting to ground, meter may be in start-up.

Caution:

If the measuring tube/s fail, process fluid can leak into the outer casing. If this happens,
depressurise the meter and remove it from the process line as soon as it safe to do so.

OPTIMASS 1000

Resistance (Ohm)

Driver

Sensor A/B

Size 15

240

78

Size 25

240

64

Size 40

168

78

Size 50

168

64

Size 15 – Ex

240

78

Size 25 – Ex

240

64

Size 40 – Ex

91

78

Size 50 – Ex

91

64

OPTIMASS

Inductance (mH)

Resistance (Ohm)

Driver

Sensor A/B

Driver

Sensor A/B

8000

2.2

0.735

38

12.5

9000

2.6

0.95

67

25