KROHNE OPTIMASS with MFC 300 EN User Manual
Page 82

82
Terminals
I/Os
CG-No
D-
D
C-
C
B-
B
A-
A
A+
Modular
Option
4
_
_
Pa / Sa
(changeable)
Ia + HART®
active
max. 2 option modules for
term. B + A: Ia or Pa / Sa or Ca
8
_
_
Pa / Sa
(changeable)
Ip + HART®
passive
max. 2 option modules for
term. B + A: Ia or Pa / Sa or Ca
6
_
_
PP / Sp
(changeable)
Ia + HART®
active
max. 2 option modules for
term. B + A:
Ia or Pp / Sp or Cp
B
_
_
PN / SN
(changeable)
Ia + HART®
active
max. 2 option modules for
term. B + A: Ip or Pp / Sp or Cp
7
_
_
PN / SN NAMUR
(changeable)
Ia + HART®
active
max. 2 option modules for
term. B + A: Ia or PN / SN or CN
C
_
_
PN / SN NAMUR
(changeable)
Ip + HART®
passive
max. 2 option modules for
term. B + A: Ip or PN / SN or CN
PA - Bus
PROFIBUS
option
D
_
_
Term
PA-
Term
PA+
Term
PA-
Term
PA+
max. 2 option modules for
term. B + A: Ia or Pa / Sa or Cp
FF - Bus
Foundation
Field-Bus
Option
E
_
_
Term
V/D-
Term
V/D+
Term
V/D-
Term
V/D+
max. 2 option modules for
term. B + A: Ia or Pa / Sa or Cp
DP - Bus
PROFIBUS
Option
F
_
0
RxD/TxD
N
RxD/TxD
P
Termin.
N
RxD/TxD
N
RxD/TxD
P
Termin.
P
max. 1 option module
for term. A: see table
below for selection
RS485
MODBUS
1
G
-
-
RS485
Many combinations possible
RS485
MODBUS
2
H
-
-
RS485
Many combinations possible
Option Modules
Abbreviation
Description
Ident for CG No.
Ia
Active current output
A
Ip
Passive current output
B
Pa / Sa
Active pulse, frequency, status output or limit switch
C
Pp / Sp
Passive pulse, frequency, status output or limit switch
E
PN / SN
Pulse, frequency, status output or limit switch to NAMUR
F
Ca
Active control input
G
Cp
Passive control input
K
CN
Control input to NAMUR
H
-
No module installed
8
-
No further module possible
0
1
Without termination resistor
2
With termination resistor