106

C Setup level

Ref

Display

Description & Settings

C1

Process Input

C1.1

Calibration

C1.1.1

Zero Calibration

Display of current zero value, continue with Query: calibrate zero? Select
with

or

:

Break: return with

g

,

Automatic:

continue with

g

, time counts down, displays actual value

when complete

Default:press

g

to set to factory-set zero value

Manual:continue with

g

display of last set value, use

to set new

value (range -10... +10%)

(preferably use "Automatic“,Before calibration, set "zero“ flow in the
pipeline!)

C1.1.2

Zero Add Offset

Direct input of a zero flow offset

C1.1.3

Pipe Diameter

Set the pipe diameter in mm for the velocity calculation

C1.1.4

Flow Correction

Defines additional correction for the mass flow range (-100...100%)

C1.2

Density

C1.2.1

Density

Perform density calibration see section 10.5

C1.2.2

Density Mode Sel

Select From:

Actual:

return with

g

Fixed:

Use a fixed value for the density

Referred:

Calculates the process density to a reference temperature

C1.2.3

Fixed Density Value

Set the fixed density value

C1.2.3

Density Reference
Temperature

Set the reference temperature for the referred density option

C1.2.4

Referred Density
Slope

Set the density slope for the referred density option

C1.3

Filter

C1.3.1

Flow Direction

Define normal direction of flow. Set either POSITIVE or NEGATIVE accord-
ing to flow arrow

C1.3.2

Pressure Suppression
Time

Define pressure suppression time limit. Range 0.0...20.0 Seconds

C1.3.3

Pressure Suppression
Cutoff

Define pressure suppression cutoff value. Range 0.0...10.0%

C1.3.4

Density Averaging

Define time constant for density measurement. Range 1.0...20.0 Seconds

C1.3.5

Low Flow Cut-Off

Define low flow cut-off value Range 00.0...10.0%

C1.4

System Control

C1.4.1

Sys Ctrl
Function

Defines action of the process control. Select

NO ACTION: OFF

Flow and Total = 0: Force flow and all totalisers to zero

Flow = 0: Force flow to zero

C1.4.2

Sys Ctrl
Condition

Defines the condition for the process control function. Select either
DENSITY or TEMPERATURE