C setup level – KROHNE OPTIMASS with MFC 300 EN User Manual
Page 106
106
C Setup level
Ref
Display
Description & Settings
C1
Process Input
C1.1
Calibration
C1.1.1
Zero Calibration
Display of current zero value, continue with Query: calibrate zero? Select
with
or
:
Break: return with
g
,
Automatic:
continue with
g
, time counts down, displays actual value
when complete
Default:press
g
to set to factory-set zero value
Manual:continue with
g
display of last set value, use
to set new
value (range -10... +10%)
(preferably use "Automatic“,Before calibration, set "zero“ flow in the
pipeline!)
C1.1.2
Zero Add Offset
Direct input of a zero flow offset
C1.1.3
Pipe Diameter
Set the pipe diameter in mm for the velocity calculation
C1.1.4
Flow Correction
Defines additional correction for the mass flow range (-100...100%)
C1.2
Density
C1.2.1
Density
Perform density calibration see section 10.5
C1.2.2
Density Mode Sel
Select From:
Actual:
return with
g
Fixed:
Use a fixed value for the density
Referred:
Calculates the process density to a reference temperature
C1.2.3
Fixed Density Value
Set the fixed density value
C1.2.3
Density Reference
Temperature
Set the reference temperature for the referred density option
C1.2.4
Referred Density
Slope
Set the density slope for the referred density option
C1.3
Filter
C1.3.1
Flow Direction
Define normal direction of flow. Set either POSITIVE or NEGATIVE accord-
ing to flow arrow
C1.3.2
Pressure Suppression
Time
Define pressure suppression time limit. Range 0.0...20.0 Seconds
C1.3.3
Pressure Suppression
Cutoff
Define pressure suppression cutoff value. Range 0.0...10.0%
C1.3.4
Density Averaging
Define time constant for density measurement. Range 1.0...20.0 Seconds
C1.3.5
Low Flow Cut-Off
Define low flow cut-off value Range 00.0...10.0%
C1.4
System Control
C1.4.1
Sys Ctrl
Function
Defines action of the process control. Select
•
NO ACTION: OFF
•
Flow and Total = 0: Force flow and all totalisers to zero
•
Flow = 0: Force flow to zero
C1.4.2
Sys Ctrl
Condition
Defines the condition for the process control function. Select either
DENSITY or TEMPERATURE