1 Point Calibration
The options are: Empty, Pure water, Town water and Other. Select with
or
and press
g
If
you select “Other” you will need to enter the product density and this can be in any of the nor-
mal density units. If you select: ‘pure water’, ‘air’ or ‘town water’ the density does not need to
be entered.
Once selected
Single Pt Density Calib
Break
is displayed.
Press
or
to select OK . Density calibration should take about 10 seconds. After this time
the result of the calibration will be displayed. CALIB OK - the point has been entered correctly.
Density Calibration
Menu C1.2.1
A density calibration can be made on-site to improve density accuracy. The original factory cali-
bration can also be re-loaded in the event of a data loss or error in density calibration proce-
dure.
Options available:
Option
Result
1 Point Calibration:
The converter uses the existing calibration, and decides which
point to adjust when the calibration is made
2 Point Calibration:
The user enters the two points to be used in the calibration.
Default
Converter restores the factory density calibration
Manual
User can read the existing density calibration values and edit if
required
Break
OK
Key
Text Displayed
Description and Settings
>
A
Quick Setup
Countdown from 2.5 s to 0.0 s, then release key
x 2
C
Setup
>
x 2
Calibration
Density
>
x2
Density Calibration?
Break
Press
g
to exit from density calibration
Density Calibration?
default
Press
g
to load factory density calibration
Density Calibration?
Manual
Press
g
to read and edit existing calibration
Density Calibration?
2 Point Calibration
Press
g
to start 2 point calibration
Density Calibration?
1 Point Calibration
Press
g
to start 1 point calibration
Default
Manual
2 Point Calibration
1 Point Calibration
Break