KROHNE OPTIMASS with MFC 300 EN User Manual

Page 121

background image

121

1 Point Calibration

The options are: Empty, Pure water, Town water and Other. Select with

or

and press

g

If

you select “Other” you will need to enter the product density and this can be in any of the nor-
mal density units. If you select: ‘pure water’, ‘air’ or ‘town water’ the density does not need to
be entered.

Once selected

Single Pt Density Calib
Break

is displayed.

Press

or

to select OK . Density calibration should take about 10 seconds. After this time

the result of the calibration will be displayed. CALIB OK - the point has been entered correctly.

Density Calibration

Menu C1.2.1

A density calibration can be made on-site to improve density accuracy. The original factory cali-
bration can also be re-loaded in the event of a data loss or error in density calibration proce-
dure.

Options available:

Option

Result

1 Point Calibration:

The converter uses the existing calibration, and decides which
point to adjust when the calibration is made

2 Point Calibration:

The user enters the two points to be used in the calibration.

Default

Converter restores the factory density calibration

Manual

User can read the existing density calibration values and edit if
required

Break

OK

Key

Text Displayed

Description and Settings

>

A

Quick Setup

Countdown from 2.5 s to 0.0 s, then release key

x 2

C

Setup

>

x 2

Calibration

Density

>

x2

Density Calibration?
Break

Press

g

to exit from density calibration

Density Calibration?
default

Press

g

to load factory density calibration

Density Calibration?
Manual

Press

g

to read and edit existing calibration

Density Calibration?
2 Point Calibration

Press

g

to start 2 point calibration

Density Calibration?
1 Point Calibration

Press

g

to start 1 point calibration

Default

Manual

2 Point Calibration

1 Point Calibration

Break