KROHNE OPTIMASS with MFC 300 EN User Manual
Page 111
111
Ref
Display
Description & Settings
C 3.x.2
Pulse width
Appears only when "fixed“ activated in Fct. C 2.x.1 Pulse Shape, see
above
xxx.xx ms
setting range: 0.05…2000 ms)
(Note: max. setting value Tp [ms]
a
500.00 / (max.
pulse rate [1/s]
C 3.x.3
Max pulse rate
xxxxx.x Hz
(setting range 00000.0…10000.0 Hz, max. 120%)
when limited to 100% pulse rate
a
100 Hz:
Imax
a
100 mA
when limited to 100% pulse rate
>
100 Hz:
Imax
a
20 mA
C 3.x.4
Measurement
•
Volume total
•
Mass total
•
Concentration total 1
•
Concentration total 2
C 3.x.5
Pulse value unit
Selection of unit from one of the lists, Dependent on measured vari-
able
C 3.x.6
Pulse p.value
C 3.x.7
Polarity
•
both polarities
Plus and minus values are used
•
positive polarity
Negative values are set to 0%
•
negative polarity
Positive values are set to 0%
•
absolute
Absolute value is used
Choice of meas. value polarity, note flow direction, see C 1.5. 1
C 3.x.8
Low flow cutoff
xxxx.x … ±xxxx.x unit dependent on meas. variable selected1st value
A
2nd value (hysteresis), values around “0“ are set to “0“
C 3.x.9
Time constant
xxx.x s (setting range 000.0…100.0 s)
C 3.x.10
Invert signal
Select:
Off
switch closes on every pulse, normally open
On
switch opens on every pulse, normally closed
C 3.x.11
Special function
This function is only at output B in devices with 2 frequency outputs
connected to terminals B + D or B + A See Fct. 3.x.11 below
Select:
OFF:
switched off
Automatic range:
switched on - for indication
Activate status output accordingly
External range:
switched on- for external range change
Activate control input accordingly
C 3.x.11
Phase shift
This function is only at output B in devices with 2 frequency outputs
connected to terminals B + D or B + A See Fct. 3.x.11 below
Select:
Off
no phase shift
0° shift
signal inversion possible
90° shift
signal inversion possible
180° shift
signal inversion possible
When Fct. C 3.5.6 Polarity is set to "both polarities“, the flow direction
is indicated (e.g. +90° or -90°)
C 3.x.12
Information
Displays: Serial No, Software No, & calibration date of the circuit board
xxx.xxx - set for volume or mass per pulse
The lowest settable pulse value is calculated using the formula:
measuring range (in l/s or kg/s)
1
100% pulse rate (in 1/s)
2
1
vol. or mass, see Fct. C 2.x.06 for current output
2
see Fct. C 2.x.3 for pulse output