KROHNE OPTIMASS with MFC 300 EN User Manual

Page 111

background image

111

Ref

Display

Description & Settings

C 3.x.2

Pulse width

Appears only when "fixed“ activated in Fct. C 2.x.1 Pulse Shape, see
above

xxx.xx ms

setting range: 0.05…2000 ms)
(Note: max. setting value Tp [ms]

a

500.00 / (max.

pulse rate [1/s]

C 3.x.3

Max pulse rate

xxxxx.x Hz

(setting range 00000.0…10000.0 Hz, max. 120%)

when limited to 100% pulse rate

a

100 Hz:

Imax

a

100 mA

when limited to 100% pulse rate

>

100 Hz:

Imax

a

20 mA

C 3.x.4

Measurement

Volume total

Mass total

Concentration total 1

Concentration total 2

C 3.x.5

Pulse value unit

Selection of unit from one of the lists, Dependent on measured vari-
able

C 3.x.6

Pulse p.value

C 3.x.7

Polarity

both polarities

Plus and minus values are used

positive polarity

Negative values are set to 0%

negative polarity

Positive values are set to 0%

absolute

Absolute value is used

Choice of meas. value polarity, note flow direction, see C 1.5. 1

C 3.x.8

Low flow cutoff

xxxx.x … ±xxxx.x unit dependent on meas. variable selected1st value

A

2nd value (hysteresis), values around “0“ are set to “0“

C 3.x.9

Time constant

xxx.x s (setting range 000.0…100.0 s)

C 3.x.10

Invert signal

Select:

Off

switch closes on every pulse, normally open

On

switch opens on every pulse, normally closed

C 3.x.11

Special function

This function is only at output B in devices with 2 frequency outputs
connected to terminals B + D or B + A See Fct. 3.x.11 below

Select:

OFF:

switched off

Automatic range:

switched on - for indication
Activate status output accordingly

External range:

switched on- for external range change
Activate control input accordingly

C 3.x.11

Phase shift

This function is only at output B in devices with 2 frequency outputs
connected to terminals B + D or B + A See Fct. 3.x.11 below

Select:

Off

no phase shift

0° shift

signal inversion possible

90° shift

signal inversion possible

180° shift

signal inversion possible

When Fct. C 3.5.6 Polarity is set to "both polarities“, the flow direction
is indicated (e.g. +90° or -90°)

C 3.x.12

Information

Displays: Serial No, Software No, & calibration date of the circuit board

xxx.xxx - set for volume or mass per pulse

The lowest settable pulse value is calculated using the formula:

measuring range (in l/s or kg/s)

1

100% pulse rate (in 1/s)

2

1

vol. or mass, see Fct. C 2.x.06 for current output

2

see Fct. C 2.x.3 for pulse output