KROHNE OPTIMASS with MFC 300 EN User Manual

Page 141

background image

141

Display Messages

Description

Possible causes, remedial action

Status: C _ _ _ _
see texts below

Output values in part simulated or fixed

Checks in progress

Testing mode by invoking test or simu-
lation functions,

All or some outputs not available,

Measured values might be simulated

Message via HART or FDT, as the case may
be. Indic. via the display when outputs held
by control input or have been set to zero

Test XXXXX
see texts below

Test of relevant unit activated

Standby Mode

Meter is in standby mode

Check Control input settings.

Check A7

Status: I _ _ _ _
see texts below

Information (current measurement OK)

Counter 1/2/3
stopped

Counter 1/2/3 has stopped counting

If counter to continue counting, activate
“yes“ in Fct. C 2.y.09 Start Counter

Power Fail

Indicates that the device was out of
service for an unknown period due to a
power failure

Cause: temporary line failure, during which
counters stopped counting.

Control input A
active

Message appears when the control
input at terminals A or terminals B is
active.

This message is for information only

Control input B
active

Over range Display 1 1st line on page ? and/or ? of display

limited by filter setting, indication not
correct

Menu Display Fct. C 6.3 and/or C 6.4, select
Meas. Page 1 or 2, and increase values in
Functions

C 6.z.03 Meas. Range and/or C 6.z.04
Limitation

Over range Display 2

Backplane, sensor

The sensor data on the backplane are
no use because they have been gener-
ated with an incompatible version.

Check for software upgrades

Backplane settings

The global settings on the backplane
are no use because they have been gen-
erated with an incompatible version.

Backplane
difference

The data on the backplane differ from
the data in the display. If the data can
be used, a dialogue is indicated in the
display.

Optical interface

The optical interface is in use. The keys
on the local display are inoperative.

Keys are ready to operate again 60s after
end of data transfer / removal of the opto-
coupler