B test level – KROHNE OPTIMASS with MFC 300 EN User Manual
Page 105
105
B Test level
Ref
Display
Description & Settings
B1
Simulation
B1.1
Mass Flow
Set Value:
Confirm with
g
key and set/edit value. Confirm with
g
At prompt
“Start simulation?”, select No or Yes and press
g
to start simulation
Break:
Extras Menu without simulation
B1.2
Density
As B1.1
B1.3
Temperature
B1.4
Current Output A
B1.5
Status Output B
B1.6
Status Output C
B1.7
Pulse Output D
B2
Actual Values
B2.1
Operating Hours
Displays the actual operating hours of the device. Exit using the
g
key
B2.2
Act. Mass Flow
Displays the actual unfiltered mass flow. Exit using the
g
key
B2.3
Volume Flow
Displays the actual unfiltered volume flow. Exit using the
g
key
B2.4
Velocity
Displays the actual unfiltered velocity. Exit using the
g
key
B2.5
Density
Displays the actual unfiltered density. Exit using the
g
key
B2.6
Temperature
Displays the actual unfiltered temperature. Exit using the
g
key
B2.7
Strain MT
Displays the actual value for the measuring tube strain gauge. Exit
using the
g
key
B2.8
Strain IC
Displays the actual value for the inner cylinder strain gauge. Exit using
the
g
key
B2.9
Tube Frequency
Displays the actual tube oscillation frequency. Exit using the
g
key
B2.10
Drive Level
Displays the actual drive level for the tube. Exit using the
g
key
B2.11
Sensor A Level
Displays the actual amplitude of vibration of Sensor A or B. Exit using
the
g
key
B2.12
Sensor B Level
B2.13
2 Phase Signal
Displays the actual flow noise. Exit using the
g
key
B2.14
SE PCB Temp
Displays the actual temperature of the Sensor Electronics. Exit using
the
g
key
B2.15
BE PCB Temp
Displays the actual temperature of the Converter (Back End)
Electronics. Exit using the
g
key
B2.16
Act. Operation Mode
Displays the actual operation mode of the system. Exit using the
g
key
B3
Information
B3.1
C Number
Displays the CG (identifying) number of the electronics.
B3.2
Sensor Electronics
Displays information about the sensor electronics
B3.3
SW.REV. MS
Displays information about the device incliding HART software
B3.4
SW.REV. UIS
Displays information about the user interface of the device
B3.6
Electronic Revision ER
Displays information about the electronics revision of the sensor