5.8 Purge Port Meters and Burst Disc Meters

Purge Port Options

If the purge port option was selected at time of order, then your meter will be fitted with ½”
NPT female connections – these will be clearly identified. These connections are sealed with
NPT plugs and PTFE tape.

Important:

Do not remove these plugs.

The meter is factory sealed with a dry nitrogen gas fill and any ingress of moisture will damage
the meter. The plugs should only be removed to purge the inside of the meter case of any
product if it is suspected that the primary measuring tube has failed. If it is suspected that the
primary tube has failed, depressurise and remove the meter from service, as soon as it is safe
to do so.

Burst Disc meters

OPTIMASS 2000 meters that have been ordered with a burst (rupture) disc will be suppled with
the disc fitted. The disc failure pressure is 20barg @ 20°C.

Meters fitted with connection ratings in excess of 100 barg,(1450 psig) will be supplied with a
burst disc fitted.

Important:

The burst disc is suitable for the designed application according to the process conditions and
flow rates as per original order. If conditions alter, consult KROHNE for further advice regard-
ing suitability of disc fitted.

If the product is in any way hazardous, it is strongly recommended that an exhaust tube is con-
nected to the NPT male thread of the burst disc so that the discharge can be piped to a safe
area. This tube should be large enough that pressure cannot build up in the meter case.

Ensure arrow on burst disc is pointing away from meter

5.9 Technical Data

Maximum Flow Rates

100

150

250

Kg/h

420,000

900,000

2,300,000

Lbs/min

14,698

33,804

84,510

Minimum Flow Rates

Kg/h

Dependant on measuring error required

Lbs/min

Dependant on measuring error required

Weights (PN40 Flanges)

100

150

250

kg

lbs

kg

lbs

kg

lbs

Compact with aluminium MFC 300

84.8

187

211.5

466

444.5

980

Compact with SS MFC 300

90.1

198

216.8

478

449.8

991

Remote with aluminium J box

80.8

178

207.8

457

440.5

971

Remote with SS J box

81.7

180

208.4

459

441.4

973

Cooling

Please consult KROHNE if cooling medium is to be used in the heating jacket.