117

Ref

Display

Description & Settings

C 6.7.16

Density

kg/m³

kg/l

lb/ft³

lb/gal

• SG

API

user-defined unit (free unit)

C 6.7.17

Text free unit

appear only when "free unit“ selected in Fct. C 6.7.16

To set these two functions, see "free unit“ below

C 6.7.18

[kg / m³] * Factor

Free (user-defined) unit

Set required texts:

For volume rate of flow, mass rate of flow and density:
max. 3 characters before and max. 3 characters after the slash

For volume and mass: max.3 characters

Permissible characters a…z / A…Z / 0…9 / . , “ + - ? / # @ $ % ~ ( ) [ ]

Set conversion factor:

Wanted unit = [unit, see above] × conversion factor

Conversion factor: max. 9 digits

Shift decimal point with

(to left) and with

(to right)

C 6.8

HART

C 6.8.1

HART

HART on

HART off

C 6.8.2

Address

If 0, HART current output has normal function Otherwise, set address
for multi-drop usage. Current set to 0% value.

C 6.8.3

Message

Hart message – free text

C 6.8.4

Description

Hart description – free text

C 6.9

Quick Setup

C 6.9.1

Reset Totaliser 1

yes

no

Selects whether counter can be reset in the Quick Setup Menu

C 6.9.2

Reset Totaliser 2

C 6.9.3

Reset Totaliser 3

To reset counters

Key

Text Displayed

Description

>

A

Quick Setup

Countdown from 2.5 s to 0.0 s, then release key

>

A3

Reset

>

A 3.1

Reset errors

A 3.2

Totaliser 1

Select totaliser that is to be reset

A 3.3

Totaliser 2

A 3.4

Totaliser 3

>

A 3.x

Reset Totaliser? No

( )

A 3.x

Reset Totaliser? Yes

4x

g

Measuring mode

Totaliser has been reset

To delete error messages (for list of possible error messages, see Sect. x.x)

>

A 3.x

Reset? No

( )

A 3.x

Reset? Yes

4x

g

Measuring mode

Errors have been reset