Display and operator control Type

Graphic display (backlit – white) 128 × 64 pixels / 59 × 31
mm

Display functions

4 pages (page through with

)

pages 1-2:
optionally with one to three lines.

Each line can be set to show required measurement
value.

In a 2-line setting, the measured variable in the

1st line can be shown in the form of a bar graph in
the 2nd line.

Display ranges and number of places freely selec-
table.

Page 3: List of diagnostic and status messages

Page 4: Trend display

Counter places

Max. 8

Language of
display texts

English, German, French, Spanish. Danish, Polish,
Portuguese, Dutch and others pending.

Operating
elements

4 optical keys (

>

g

) for operator control of the

signal converter without opening the housing

Infrared interface for reading and writing all parameters
with KROHNE IR-Interface without opening the housing

Output / input assemblies

For number and possible combinations of the various
outputs and inputs, refer to section. 9.7

Current outputs

Function

Mass and volume flow rate, flow velocity, tempera-
ture, concentration, concentration flow, density,
diagnosis values

HART® interface is standard (but not for all option
modules), see section. 9.8

active or passive operation, dependent on the output
/ input assemblies, see section 9.8

Op. values and load
rating

Active

I

a

22 mA / RL

a

1 k

Ω

Passive I

a

22 mA / U

a

32 V DC

Current

Measuring range Imin - Imax between 0...20 mA set-

table as required

Overrange

Setting: 0.00 mA

a

value

a

21.5 mA

Error identification

0 mA

a

IErr

<

Imin or Imax

<

IErr

a

22 mA

Forward / reverse
measurement

Direction identified via status output, see below

Automatic range or
external range

via status output or control input, see below

Time constant

0 - 100.0 s, settable as required

Low-flow cutoff

Value: 0.0...20.0 %

of Q100% settable

as required

Hysteresis: ± 0.0...20.0 %.