KROHNE OPTIMASS with MFC 300 EN User Manual
Page 96
96
Display and operator control Type
Graphic display (backlit – white) 128 × 64 pixels / 59 × 31
mm
Display functions
4 pages (page through with
)
•
pages 1-2:
optionally with one to three lines.
Each line can be set to show required measurement
value.
In a 2-line setting, the measured variable in the
1st line can be shown in the form of a bar graph in
the 2nd line.
Display ranges and number of places freely selec-
table.
•
Page 3: List of diagnostic and status messages
Page 4: Trend display
Counter places
Max. 8
Language of
display texts
English, German, French, Spanish. Danish, Polish,
Portuguese, Dutch and others pending.
Operating
elements
4 optical keys (
>
g
) for operator control of the
signal converter without opening the housing
Infrared interface for reading and writing all parameters
with KROHNE IR-Interface without opening the housing
Output / input assemblies
For number and possible combinations of the various
outputs and inputs, refer to section. 9.7
Current outputs
Function
•
Mass and volume flow rate, flow velocity, tempera-
ture, concentration, concentration flow, density,
diagnosis values
•
HART® interface is standard (but not for all option
modules), see section. 9.8
•
active or passive operation, dependent on the output
/ input assemblies, see section 9.8
Op. values and load
rating
Active
I
a
22 mA / RL
a
1 k
Ω
Passive I
a
22 mA / U
a
32 V DC
Current
Measuring range Imin - Imax between 0...20 mA set-
table as required
Overrange
Setting: 0.00 mA
a
value
a
21.5 mA
Error identification
0 mA
a
IErr
<
Imin or Imax
<
IErr
a
22 mA
Forward / reverse
measurement
Direction identified via status output, see below
Automatic range or
external range
via status output or control input, see below
Time constant
0 - 100.0 s, settable as required
Low-flow cutoff
Value: 0.0...20.0 %
of Q100% settable
as required
Hysteresis: ± 0.0...20.0 %.