5 status messages and diagnostics information – KROHNE OPTIMASS with MFC 300 EN User Manual

Page 138

background image

138

Display Messages

Description

Possible causes, remedial action

Status: F _ _ _ _
See texts below)

Operational fault in device

mA output

A

3.5 mA

Status outputs open

Pulse / frequency output: no pulses

Repair necessary!

Error in device

Operational fault in device.
No measurement possible.

Group message, when one of the following
or some other severe error occurs

IO 1

Error, operational fault in IO 1.
No measurement possible

Load settings (Fct. C 6.6.3) (Backup 1 or
Backup 2 or factory settings).

If status message continues, replace elec-
tronic unit.

IO 2

Error, operational fault in IO 2,
No measurement possible

Parameter

Error, operational fault in device,
Parameters no longer useable
No measurement possible

Configuration
(also when module
changed)

Identified configuration differs from the
one stored.
No measurement possible

After change of module, confirm query for
changed configuration.

If device configuration is unchanged the
device is likely to be defective.

Replace electronic unit.

Display

Error, operational fault in display

Defective, replace electronic unit

SE defective

Error, operational fault in sensor elec-
tronics,

No measurement possible

Defective, replace electronic unit

sensor global

Data error in the global data of the sen-
sor electronic equipment

No measurement possible

Load settings (Fct. C 6.6.3) (Backup 1 or 2
or factory settings).

If status message continues, replace elec-
tronic unit.

sensor local

Data error in the local data of the sen-
sor electronics.

No measurement possible

Defective, replace electronic unit

SE Data Error

Data error in the Sensor Electronics

Load settings (Fct. C 6.6.3) (Backup 1 or 2
or factory settings).

If status message continues, replace elec-
tronic unit.

current output
A/B/C

Operational fault in current output at
the pair of terminals indicated.
Measured value at this pair of terminals
no longer available

Defective

Replace electronic unit or I/O module.

Fieldbus

Operational fault in Fieldbus
Connections ???

Defective

Replace electronic unit or I/O module

SE Drive Failure

Fault in Sensor Electronics, unable to
control drive amplitude

Replace sensor electronics

11.5 Status Messages and Diagnostics Information

These messages are shown on page 3 of the display. mA- and status outputs can be set so that
error messages are signalled, see Section 10.4, Fct. C 3.x.4 (current) and Fct. C 3.x.1 (status).

Please note:

for settings with mA and/or status outputs:

- out of specification (S) = all errors of Categories S, F and F

- application error (F) = all errors of Categories F and F

- fault in device (F)

= all errors of Category F