5 status messages and diagnostics information – KROHNE OPTIMASS with MFC 300 EN User Manual
Page 138
138
Display Messages
Description
Possible causes, remedial action
Status: F _ _ _ _
See texts below)
Operational fault in device
•
mA output
A
3.5 mA
•
Status outputs open
•
Pulse / frequency output: no pulses
Repair necessary!
Error in device
Operational fault in device.
No measurement possible.
Group message, when one of the following
or some other severe error occurs
IO 1
Error, operational fault in IO 1.
No measurement possible
Load settings (Fct. C 6.6.3) (Backup 1 or
Backup 2 or factory settings).
If status message continues, replace elec-
tronic unit.
IO 2
Error, operational fault in IO 2,
No measurement possible
Parameter
Error, operational fault in device,
Parameters no longer useable
No measurement possible
Configuration
(also when module
changed)
Identified configuration differs from the
one stored.
No measurement possible
After change of module, confirm query for
changed configuration.
If device configuration is unchanged the
device is likely to be defective.
Replace electronic unit.
Display
Error, operational fault in display
Defective, replace electronic unit
SE defective
Error, operational fault in sensor elec-
tronics,
No measurement possible
Defective, replace electronic unit
sensor global
Data error in the global data of the sen-
sor electronic equipment
No measurement possible
Load settings (Fct. C 6.6.3) (Backup 1 or 2
or factory settings).
If status message continues, replace elec-
tronic unit.
sensor local
Data error in the local data of the sen-
sor electronics.
No measurement possible
Defective, replace electronic unit
SE Data Error
Data error in the Sensor Electronics
Load settings (Fct. C 6.6.3) (Backup 1 or 2
or factory settings).
If status message continues, replace elec-
tronic unit.
current output
A/B/C
Operational fault in current output at
the pair of terminals indicated.
Measured value at this pair of terminals
no longer available
Defective
Replace electronic unit or I/O module.
Fieldbus
Operational fault in Fieldbus
Connections ???
Defective
Replace electronic unit or I/O module
SE Drive Failure
Fault in Sensor Electronics, unable to
control drive amplitude
Replace sensor electronics
11.5 Status Messages and Diagnostics Information
These messages are shown on page 3 of the display. mA- and status outputs can be set so that
error messages are signalled, see Section 10.4, Fct. C 3.x.4 (current) and Fct. C 3.x.1 (status).
Please note:
for settings with mA and/or status outputs:
- out of specification (S) = all errors of Categories S, F and F
- application error (F) = all errors of Categories F and F
- fault in device (F)
= all errors of Category F