beautypg.com

KROHNE OPTIMASS with MFC 300 EN User Manual

Page 51

background image

51

Size

Max Force: Flanges

Max Force: Hygienic

Connectors

06 S

19 kN

1.5 kN

10 H/S

25 kN

2 kN

15 H/S

*

38 kN

5 kN

25 H/S

60 kN

9 kN

40 H/S

80 kN

12 kN

50 H/S

80 kN

12 kN

80 H/S

170 kN

18 kN

*

On OPTIMASS 15H or S with ½” ASME flanges only, the max. end load is 19kN

Hastelloy and Stainless Steel

Maximum Pipework Forces

The maximum forces exerted on the meter from the pipe work, compressive or tensile has
been calculated for the 7000 Series (Straight tube meter) with Titanium, Hastelloy and SS
measuring tubes as follows:

Loads given in both tables are maximum static loads. If loads are cycling, particularly between
tension and compression then these loads should be reduced.

For information regarding Tantalum (or any material) please consult KROHNE.

Size

Max Force: Flanges

Max Force: Hygienic

Connectors

06 T

19 kN

1.5 kN

10 T

25 kN

2 kN

15 T

*

38 kN

5 kN

25 T

60 kN

9 kN

40 T

80 kN

12 kN

50 T

170 kN

12 kN

80 T

230 kN

30 kN

*

On OPTIMASS 15T with ½” ASME flanges only, the max. end load is 19kN

Titanium