11.1 Diagnostic functions

The following diagnostic functions are available in the TEST level Menu B2.

Temperature (menu B2.6):

Displays temperature in either oC or oF. The value should be stable.

Strain (menu B2.7 Strain measuring tube / B2.8 Strain inner cylinder):

Value of strain in Ohms. The values should be in the range stated in section 10.3. Wildly
unstable value even after temperature stabilisation: the strain gauge has possibly become
delaminated due to the meter being operated over maximum temperature for prolonged
periods of time (please contact KROHNE UK service department).

Frequency (menu B2.9):

Variations in the first digit after the decimal point indicate gas or air in the fluid.

Worn or eroded flow tube: frequency will increase by around 2...4 Hz meter requires re-cal-
ibration

Coatings can also alter the frequency

Large fluctuations are seen if the meter is in ‘Start Up’

Drive Energy (menu B2.10):

Typical values for the drive energy level with water as process fluid (with no air or gas) are:

Higher drive energy levels can occur due to gas or air in the fluid or at the measurement of
high viscous fluids or fluids with high densities.

Sensor A and B (menu B2.11, B2.12):

The displayed value should be around:

80% for OPTIMASS 7000 – sizes 06 … 40, OPTIMASS 8000, OPTIMASS 9000

60% for OPTIMASS 7000 – sizes 50 and 80

55% for OPTIMASS 3000 - all sizes

60% for OPTIMASS 2000 - size 100

50% for OPTIMASS 2000 - size 150 & 250

80% for OPTIMASS 1000

NOTE: Sensor values should be within 2% of each other.

2-Phase Flow (menu B2.13):

This function gives an indication of the 2-phase signal level of the unit. This is application and
process dependent and it can be used to determine the set point for the 2-phase flow alarm
function. This can only be done on-site under actual flowing conditions.

SE Board Temperature (menu B2.14):

Displays the temperature of the sensor electronics

BE Board Temperature (menu B2.15)):

Displays the temperature of the converter electronics

OPTIMASS

11

SERVICE & TROUBLE SHOOTING

OPTIMASS 1000:

All sizes

0...6

OPTIMASS 2000:

All sizes

0...5

OPTIMASS 3000:

all sizes:

0...4

OPTIMASS 7000 :

06...40

50...80

0...6

4...10

OPTIMASS 8000:

all sizes:

0...5

OPTIMASS 9000:

all sizes:

0...5