beautypg.com

4 changing display oriention – KROHNE OPTIMASS with MFC 300 EN User Manual

Page 74

background image

74

1

Cover, electronics compartment

2

Cable entry for inputs/outputs

3

Cable entry for power

4

Cover for terminal compartment for power supply and inputs/outputs

F Version only (remote)

5

Attachment plate for wall or pipe mounting

6

Locking screw for sensor terminal compartment cover

7

Cable entry for sensor cable

8

Sensor terminal compartment cover

9.4 Changing Display Oriention

The orientation of the meter, the display of the MFC 300 C and MFC 300 F can be rotated in 90°
steps.

1.

Unscrew the electronics compartment front cover.

2

Pinch the two plastic locking clips either side of the display to release it.

3.

Rotate the display to the required position and push back onto the clips until they lock.

Note Make sure not to damage the flat ribbon cable!

4.

Replace the cover and tighten by hand.

Important:

Before replacing the electronics compartment cover, ensure that the threads on the converter
housing are clean of debris and apply grease to the cover threads prior to re-fitting.

This is particularly important with hazardous-duty (Ex) versions.

Wall:

1. Using the Dimensions in section 9.10, drill the wall and prepare with rawplugs

2. Secure with fixing screws and lock washers.

1

3

4

7

2

5

8

6