Low Flow Threshold

3 4

Pressure Supression

Settings of these parameters depend on actual process conditions and characteristics of the
pipework and so can only be determined by experimentation in-situ.

Process Control (Menu C1.4.1 - Function)

This menu allows the setting up of certain instrument functions depending on the selected
process condition. If a pre-determined condition arises (as selected in Fct. 4.8.2) then one of
the following options can be selected. Options are:

NO ACTION : Process control is OFF

Flow and Total = 0 : Forces flow and counters to zero

Flow = 0 : Force flow to zero

Menu C1.4.2 - Condition

Selects the process condition that activates the process control function. Select either Density
or Temperature.

Menu C1.4.3 – Max Limit

Menu C1.4.4 – Min Limit

Sets the limits that activate the process control function.

Values outside of these limits activate the function.

Diagnosis Values (Menu C1.5.4 to C1.5.6)

Selects the values for the diagnosis values. These can be then attributed to the display or out-
puts.

Graph

hic Page (Menu C6.5)

With the MFC300, it is possible to show a graphical trend of the prime measured value which is
defined by the first measured value on the first measuring page.

Menu C6.5.1 defines the range of the trend value (manual or automatic)

Menu C6.5.2 defines the manual range

Menu C6.5.3 defines the time span of the trend