6 hygienic and sanitary connections (all sizes) – KROHNE OPTIMASS with MFC 300 EN User Manual
Page 66
66
8.6 Hygienic and Sanitary Connections (all sizes)
Maximum pressure: 10 barg at 150°C or 145 psig at 302°F
Maximum pipe work forces
Forces exerted on the meter from the process pipe are not permitted. Mechanical installation
should be designed to prevent such forces.
Hygienic Applications
The OPTIMASS 8000/9000 series is available with a variety of hygienic process connectors.
When using / installing meters with hygienic process connectors, care should be taken to
ensure the meter is well supported / clamped, as the meters are heavy and could injure when
disconnected from the adjacent pipe work.
The recommended method of installation is to mount the meter against a support / wall with
the body of the meter supported/clamped. The process pipe work can then be supported off the
meter.
The meter is too heavy to be supported from the thin walled piping usually associated with the
hygienic industry.
1 2
Support Points
Installation Lengths:
For installation lengths, please see section 8.9.
Please check with KROHNE if you are unsure of the installation length. Many meters are built
to customer requirements / specifications especially where special hygienic process connec-
tors have been adapted to the meter. As these are normally non-standard, the installation
length will not be given in the technical data.
It is also recommended that the seals be replaced regularly to maintain the hygienic integrity
of the connection.
Hygienic Connection Materials
Material: SS 316L
Unless specifically requested, internal surfaces are not polished and no warranty is made as to
the surface finish.
1
2