9.7.3 Modular I/Os (input/output versions)
•
The grey boxes denote freely selectable option modules for terminals A and B.
•
Terminal A+ functions only for the Basic I/O.
•
For hazardous areas, all I/O variants for MFC 300 C and MFC 300 F are available with ter-
minal compartment in EEx - d (flameproof enclosure) or EEx - e (increased safety) protec-
tion.
9.7.1 Basic I/Os (input/output versions) – not changeable
Terminals
I/Os
CG-No
D-
D
C-
C
B-
B
A-
A
A+
Basic
Standard
Pp / Sp
(changeable)
Sp
Sp / Cp
(changeable)
Ip + HART®
1
0
0
or (reverse term.)
Ia + HART®
EEx - i
Option
2
0
0
PN / SN NAMUR
(changeable)
Ia + HART®
active
3
0
0
PN / SN NAMUR
(changeable)
Ip + HART®
passive
2
1
0
PN / SN NAMUR
(changeable)
Ia + HART®
active
PN / SN / CN
NAMUR
(changeable)
Ia
3
1
0
PN / SN NAMUR
(changeable)
Ip + HART®
active
PN/SN/CN NAMUR
(changeable)
2
2
0
PN / SN NAMUR
(changeable)
Ia + HART®
active
PN/SN/CN NAMUR
(changeable)
Ip
3
2
0
PN / SN NAMUR
(changeable)
Ip + HART®
passive
PN / SN / CN
NAMUR
(changeable)
Ip
PA - Bus
PROFIBUS
(EEx-i)
option
D
0
0
Term PA- Term PA+ Term PA- Term PA+
FISCO Device
FISCO Device
D
1
0
Term PA- Term PA+ Term PA- Term PA+
PN / SN / CN
NAMUR
(changeable)
Ia
FISCO Device
FISCO Device
D
2
0
Term PA- Term PA+ Term PA- Term PA+
PN / SN / CN
NAMUR
(changeable)
Ip
FISCO Device
FISCO Device
FF - Bus
Foundation
Field-Bus
(EEx-i)
Option
E
0
0
Term V/D- Term V/D+ Term V/D- Term V/D+
FISCO Device
FISCO Device
E
1
0
Term V/D- Term V/D+ Term V/D- Term V/D+
PN / SN / CN
NAMUR
(changeable)
Ia
FISCO Device
FISCO Device
E
2
0
Term V/D- Term V/D+ Term V/D- Term V/D+
PN / SN / CN
NAMUR
(changeable)
Ip
FISCO Device
FISCO Device
9.7.2 Fixed I/Os (input/output versions) not changeable