Page 113

background image

113

Ref

Display

Description & Settings

C 3.x

Control input X

C 3.x.1

Mode

off

zero output + stop cnt.
(not display)

stop all counters

stop counter 1, 2 or 3

reset all counters

reset counter 1, 2 or 3

error reset

hold all outputs
(not display, not counters)

hold output W, Y or Z

all outputs zero
(not display, not counters)

output W, Y or Z zero

range change W, Y or Z

zero calibration

Note! If two control inputs are provided, they should not be set to the
same operating mode; if they are, only the control input connected to
Term. A is in function!

C 3.x.2

Invert signal

off

High current at input, switch closed activates function

on

Low current at output, switch open activates function

C 3.x.3

Information

Displays the Serial No, Software No, and calibration date of the circuit
board

C 3.x.4

Simulation

For sequence, see Test Level Menu, Fct. B 1

C4 I/O Totaliser

C 4.y

Totaliser 1, 2 or 3

All functions and settings for both counters are the same! In the fol-
lowing descriptions the “y“ indicates the counter:
Totaliser 1 = C 3.1

Totaliser 2 = C 3.2

Totaliser 3 = C 3.3

C 4.y.1

Function of Totaliser

Incremental Total

counts only positive values

Decremental Total

counts only negative values

Absolute Total

counts positive and negative values

off

counter is disabled

C 4.y.2

Measurement

volume flow

mass flow

concentration total1

concentration total2

C 4.y.3

Low flow cutoff

xxxx.x … ±xxxx.x unit dependent on measured variable selected
1st value A 2nd value (hysteresis), values around “0“ are set to “0“

C 4.y.4

Time constant

xx.x s

(setting range 000.0…100.0 s)

C 4.y.5

Preset value

x.xxxxx in the unit selected, max. 8 places (see Fct. C 4.7.10 or 13),
Status output X becomes active when value is reached. Status output
X mode (see Fct. C 3.x.1) must be set to Counter 1/2/3 Preset

C 4.y.6

Reset Total

yes

no

C 4.y.7

Set Total

Set initial counter value (overwrites current reading)

Cancel

g

>

return without simulation

Set Value

g

>

set value

g

>

query “set counter?“ No - Yes

>

execute with

g

C 4.y.8

Stop Totaliser

Select:

yes

no

C 4.y.9

Start Totaliser

Select:

yes

no

C 4.y.10

Information

Displays the Serial No, Software Number and calibration date of the
circuit board.