Q1. The beep keeps sounding.
A1. The beep sounds according to the pass/fail test result, which in
this case is set to repetitive mode. Do one of the following.
•
Set the test mode to Single (manual trigger), so that the beep
sounds only when the test is initiated manually. Press the
Sing/Rep key to change the setting. For details, see page55.
•
Turn Off the beep entirely. Press the Menu key, F5 (System),
move the cursor to Beep and select Off using the arrow keys. For
details, see page61.
Q2. Panel operation seems disabled.
A2. Panel (local) operation is disabled in the Remote control mode
(page107). Press the Local key to bring back the instrument to local
operation mode (remote control is cancelled).
Q3. I cannot see the display clearly.
A3. Use the display contrast knob on the rear panel to adjust the
brightness.
For more information, contact your local dealer or GWInstek at
www.gwinstek.com
LCR-8000G Series User Manual
130
130
130
130
A
PPENDIX
Fuse Replacement
Step
1. Take off the power cord and remove the fuse
socket using a minus driver.
AC
115
/ 230
V
50 /
60H
z
FUS
E 2
50V
T 3
A
2. Replace the fuse in the holder.
Rating
T3A/250V