Emote control, Interface configuration – GW Instek LCR-8000G Series User Manual
Page 54
REMOTE CONTROL
107
107
107
107
R
EMOTE CONTROL
This chapter describes basic aspects of IEEE488.2
based remote control. Both RS-232C and GPIB
interface can be used for remote control.
Interface
Configuration
Configure RS-232C interface .......................... 108
Configure GPIB interface ............................... 109
Command syntax Command Syntax .......................................... 111
Command set
System command ..................................................... 112
Measurement command .......................................... 113
Multi-step program command................................. 115
Calibration command ............................................... 117
Graph command....................................................... 118
LCR-8000G Series User Manual
108
108
108
108
Interface Configuration
Configure RS-232C interface
Connector
DB-9, Male
RS-232C
configuration
Baud rate
9600
Parity
None
Data bit
8
Stop bit
1
Connect the RS-232C cable to
the rear panel port: DB-9 male
connector.
RS232
Pin assignment
1 2 3 4 5
6 7 8 9
2: RxD (Receive data)
3: TxD (Transmit data)
5: GND
7: RTS (Request to send)
8: CTS (Clear to send)
1, 4, 6, 9: No connection
PC connection
Use the Null Modem connection as in the below
diagram.
LCR-8101G
PC
TxD
RxD
GND
Pin2
Pin3
Pin5
TxD
RxD
GND
Pin2
Pin3
Pin5
CTS
RTS
Pin7
Pin8
CTS
RTS Pin7
Pin8