beautypg.com

Speed / step setting – GW Instek LCR-8000G Series User Manual

Page 51

background image

GRAPH MODE

101

101

101

101

Speed / Step Setting

Select measurement speed (capture timing)

Background

The speed is the same setting used in the basic
measurement (page56), except in the graph mode,
maximum speed setting is not available.

Panel operation

1. Press the Up/Down key and move

the cursor to Speed.

Speed:

Fast

2. If necessary, press the Left/Right

key to change the setting (time per
capture).

DC

AC≤
100Hz

AC≤
2kHz

AC>
2kHz

AC≥
1MHz

Slow

900ms 1.3s

600ms 600ms 620ms

Med

120ms 1.2s

470ms 450ms 470ms

Fast

60ms

650ms 180ms 150ms 150ms

LCR-8000G Series User Manual

102

102

102

102

Select step size

Background

The step size selects whether to plot every
captured data (step size 1) or to plot only the
selected data (step size 2, 4, 8 = every 2, 4, 8 data).
Step size 1: detailed graph, slow capturing
Step size 2, 4, 8: simplified graph, fast capturing

Panel operation

1. Press the Up/Down key and move

the cursor to Step size.

Step Size:

2 4 8

1

2. If necessary, press the Left/Right

key to change the setting.

Range

1 (plot all data), 2, 4, 8