CALIBRATION

127

127

127

127

5. The trimming starts and ends automatically.

SHORT CIRCUIT TRIM

IN PROGRESS

ALL FREQUENCY

Trim pass

The display goes back to the
calibration mode menu.

Trim fail

The Fail message appears on the
display. Press any other key to go
back to the original menu.

Press any key to continue...

FAILED

SHORT CIRCUIT TRIM

6. Switch the fixture setting from O/C to S/C (or

from S/C to O/C) and repeat from step 1.

Note

When an “O/S Trim Error!” message appears, it
means either that you haven’t properly performed an
O/S trimming yet or that there is not an exactly
matched calibration factor that can be used to
calculate the most precise measurement reading
based on the spot frequency you are using. When an
“O/S Trimming Error!” message appears, the LCR-
8000G will use an existing calibration factor at the
closest frequency point to the spot frequency you are

LCR-8000G Series User Manual

128

128

128

128

using to calculate a measurement. However, this
measurement is only an approximation. To obtain the
most optimized measurement accuracy and to clear
the warning message, it is best to perform O/S
trimming again, based on the actual measuring
frequency being used.

To get the most accurate measurement reading for the
frequency spot you are currently using, any of the O/S

trimming options (All-frequency, ≦10kHz, ≦100kHz

and Spot frequency) can be used despite the differing
number of total calibrated frequency points, see O/S
Trimming Points section in the Appendix for details on
page 131.