Multi-step program run – GW Instek LCR-8000G Series User Manual
Page 40

PASS-FAIL MODE
79
79
79
79
Copy (duplicate) program step
Background
Copying the step inserts a new, identical step next
to the current step (= the step where the cursor
resides).
Panel operation
Press F2 (Copy). A new step with
identical contents appears on the right
side.
F 2
Before (step 3 empty)
After (step 2 copied to 3)
Delete program step
Background
Deleting the step deleted the currently selected
step (= where the cursor resides). The other step
numbers are decreased by 1 (shifted to the left in
the table).
Panel operation
Press F3 (Delete). The current step is
deleted, and the whole table shifts to
the left.
F 3
Before (step 2 deleted)
After (step 3 becomes 2)
LCR-8000G Series User Manual
80
80
80
80
Multi-Step Program Run
Run program
Panel operation
1. When editing is completed, press F6
(Run) to run the multi-step program.
The display changes to program
running mode.
F 6
MULTI STEP MODE
F 1
PROGRAM: Demo
Run
–
Start
F 6
SET
2. Press the Sing/Rep key to select
Single (manual trigger) or Repetitive
(automatic trigger).
Sing/Rep
Manual
trigger
Manual trigger
Starts running the program
when the Trigger key or F1
(Start) is pressed.
Trig
F 1
Auto
trigger
Auto trigger
Starts running the program when the
LCR-8000G detects the DUT is
connected to the fixture (scans the
fixture continuously). Manual