GRAPH MODE

97

97

97

97

If a wrong unit is entered, the value is cancelled.

Unit Mismatched

If the entered Hi level is lower than the Lo level,

the two values are swapped.

Hi and Lo Swapped

Set Lo level

6. Press the Up/Down key to move the

cursor to Lo percentage and repeat
the same step.

Cs Lo:

-19.500%

Range

–1.0x10

12

(Tera) ~ 1.0x10

12

(Tera) %

Set Nominal

7. Press the Up/Down key to move the

cursor to Nominal.

Nominal:

1.0000mF

8. Use the numerical keys to enter the nominal

value on which the Hi and Lo percentages are
based.

Range

According to each measurement item
(see page37).

1.2mΩ

m

Enter

1.5kH

Enter

H

k

Backspace

All clear

Clear

If a wrong unit is entered, the value is cancelled.

Unit Mismatched

If a value outside of the range is entered, the

closest available value is automatically selected.

Nearest Available

LCR-8000G Series User Manual

98

98

98

98

Set vertical axis (Auto + Absolute)

Background

The Y (vertical) axis configuration is available for:

Manual or Automatic fit: Selects whether the
vertical range is manually set or automatically
adjusted as the graph is plotted.

Absolute or Percentage: Selects how the vertical
range is defined, as absolute values (minimum
and maximum) or percentage of the nominal
(center) value.

Panel operation

1. Press F2 to select Abs (Absolute).

Abs

%

F 2

2. Press F3 to select Auto Fit.

Auto Fit

F 3

3. Nothing new appears on the display: the LCR-

8000G automatically configures the vertical
scale according to the measured data.

Step Size:

2 4 8

1