GRAPH MODE

91

91

91

91

Unit Mismatched

If a value outside of the range is entered, the

closest available value is automatically selected.

Nearest Available

If the entered start Voltage level is higher than

the stop Voltage, the two values are swapped.

Hi and Lo Swapped

Set stop Voltage 4. Repeat the above step for the stop Voltage.

Stop:

1.00 V

Range

10mV ~ 2V (AC≤3 MHz)

10mV ~ 1V (AC>3 MHz) *1mV step

(Stop Voltage must be higher than the Start
Voltage)

Set measurement
Frequency

5. Use the Up/Down key to move the

cursor to Frequency setting (Freq).

Freq :

10.000kHz

Use the numerical keys to enter the
measurement frequency.

Range

20Hz ~ 1MHz/5MHz/10MHz

50Hz

Enter

1MHz

M

Enter

Backspace

All clear

Clear

Select Lin/Log
scale

6. Press F1 to select the horizontal

scale, Linear or Logarithmic.

Log(V)

Lin(V)

F 1

LCR-8000G Series User Manual

92

92

92

92

Set horizontal axis (Frequency)

Background

The X (horizontal) axis is selectable from Voltage
and Frequency sweep.

When Voltage sweep is selected: measurement
Frequency is fixed

When Frequency sweep is selected:
measurement Voltage is fixed

Select Frequency 1. Press the Up/Down key and move

the cursor to Sweep.

Sweep:

Drive Level

2. If necessary, press the Left/Right

key to change the sweep setting to
Frequency.

Drive Level

Frequency

Set start
Frequency

3. Press the Up/Down key and move

the cursor to Start.

Start:

:

20.000Hz

Use the numerical keys to enter the
starting Frequency.

Range

20Hz ~ 1MHz/5MHz/10MHz

(stop Frequency must be higher than the start
Frequency)

50Hz

Enter

1.2kHz

k

Enter

Backspace

All clear

Clear

If a wrong unit is entered, the value is cancelled.