beautypg.com

GW Instek LCR-8000G Series User Manual

Page 58

background image

REMOTE CONTROL

115

115

115

115

:meas:speed

:meas:speed?

Selects or returns measurement speed.

Send parameter: max, fast, med, slow

Return parameter: 0 (max), 1 (fast), 2 (med), 3
(slow)

:meas:test:ac

Selects AC measurement.

:meas:test:rdc

Selects Rdc measurement.

:meas:test?

Returns measurement type.

Parameter: 0 (AC measurement), 1 (Rdc
measurement)

:meas:trig

Triggers an AC or Rdc measurement manually.
Returns the 1

st

and 2

nd

measurement (only the 1

st

in

Rdc).

Example: –396.283E-6, 99.558 (uF/D)

:mode?

Query the currently selected operating mode.

:rep

:rep?

Enables or returns repetitive measurements when
unit is returned to local control.

Send parameter: on (repetitive), off (single shot)

Return parameter: 0 (single shot), 1 (repetitive)

Example: :rep on (repetitive mode)

:trig

Triggers a measurement in the current mode.

Multi-step program command

:multi:set

Switches to the multi-step set-up page.

:multi:del

Removes a step in the program.

Parameter: 1 ~ 30

Example: :multi:del 2 (deletes step 2)

LCR-8000G Series User Manual

116

116

116

116

:multi:delay

:multi:delay?

Sets or returns trigger delay time for currently
selected step in millisecond.

Parameter: 0ms ~ 1000ms

Example: :multi:delay 10m (10ms)

:multi:freq

:multi:freq?

Sets or returns the frequency for the currently
selected step in Hz.

Parameter: 20 ~ 1/5/10MHz

Example: :multi:freq 1e3 (1kHz)

:multi:func

:multi:func?

Sets or returns measurement type for the currently
selected step.

Send parameter: LS, LP, Q, CS, CP, D, Z, PHASE,
RS, RP, X, G, B, Y, RDC

Return parameter: 1 (Z), 2 (Ls), 3 (Lp), 4 (Cs), 5
(Cp), 7 (Y), 8 (G), 9 (P), 10 (Q), 11 (D), 12 (Rs), 13
(Rp), 14 (B), 15 (X), 16 (Rdc)

Example: :multi:func ls (Series inductance)

:multi:hi-lim

:multi:hi-lim?

Sets or returns the higher test limit of the currently
selected step.

Example: :multi:hi-lim 10 (limit to 10.0)

:multi:lev

:multi:lev?

Sets or returns the drive level for the currently
selected step in Voltage.

Parameter: 10mV ~ 2V (DC/AC≤3 MHz)

10mV ~ 1V (AC>3 MHz)

Example: :multi:lev 200m (200mV)

:multi:load

Loads an existed file to run or edit.

Example: :multi:load demo (file name demo)

:multi:lo-lim

:multi:lo-lim?

Sets or returns the lower test limit of the currently
selected step.

Example: :multi:lo-lim –5 (limit to –5)