GW Instek LCR-8000G Series User Manual

Page 39

background image

PASS-FAIL MODE

77

77

77

77

Set frequency

Move the cursor to Freq column.
Enter the frequency using the
numerical keys and unit keys.

Func
Freq

500.00

Ls

2.00 V

Volt

Range

20Hz ~ 1MHz/5MHz/10MHz

5 digit resolution

Example:

0.5kHz (500Hz)

k

Enter

Set voltage

Move the cursor to Volt column.
Enter the voltage using the
numerical keys and unit keys.

Freq
Volt

2.00 V

500.00

Bias

Range

(1mV step)

10mV ~ 2V (DC or AC≤3 MHz)

10mV ~ 1V (AC>3 MHz)

Example:
100mV

m

V/A

Enter

Select data
capture rate

Move the cursor to Spd column.
Select the capture rate by
pressing F1 (Prog) repeatedly.

Bias
Spd

MAX

1.0000H

Hi

F 1

DC

AC≤
100Hz

AC≤
2kHz

AC>
2kHz

AC≥
1MHz

Slow

900ms 1.3s

600ms 600ms 620ms

Med

120ms 1.2s

470ms 450ms 470ms

Fast

60ms

650ms 180ms 150ms 150ms

Max

30ms

600ms 120ms 75ms

120ms

LCR-8000G Series User Manual

78

78

78

78

Set Hi limit

Move the cursor to Hi column.
Enter the Hi limit using the
numerical keys and unit keys.

Spd
Hi

1.0000H

0.0000H

Lo

MAX

Range

follows the specification for each
measurement item

Example: 1.5kH
(for Ls)

Enter

H

k

Set Lo limit

Move the cursor to Lo column.
Enter the Lo limit using the
numerical keys and unit keys.

Spd
Hi

0.0000H

Lo

MAX

1.0000H

Range

follows the specification for each
measurement item

Example: 1.0kH
(for Ls)

Enter

H

k

Set trigger delay Move the cursor to Dly (Delay)

column. Enter the amount of
trigger delay time using the
numerical keys and unit keys.

Hi
Lo

10 ms

Dly

1.0000H

0.0000H

Range

0 (no delay) ~ 1000ms

Example: 10ms

m