beautypg.com

Fixture connection – GW Instek LCR-8000G Series User Manual

Page 12

background image

GETTING STARTED

23

23

23

23

Select AC mains frequency (50/60Hz)

Background

Although the LCR-8000G works under both 50 and
60Hz power frequencies, we recommend selecting
the frequency that matches the local setting to get
the best measurement precision, especially at
lower frequencies (< 100Hz).

Panel operation

1. Press the Menu key, then F5

(System). The system menu appears.

Menu

F 5

2. Press the Up/Down key and move

the cursor to Line frequency.

:

Line frequency

50Hz

3. If necessary, press the Left/Right

key to select the frequency, from 50
to 60Hz or from 60Hz to 50Hz.

:

Line frequency

60Hz

LCR-8000G Series User Manual

24

24

24

24

Fixture Connection

Fixture structure

Background

The standard fixture is a four-wire type with a
common terminal for screen connection. The outer
terminals (Hforce and Lforce) provide the current
and the inner terminals (Hsense and Lsense)
measures the potential.

Diagram

L

FORCE

L

SENSE

H

SENSE

H

FORCE

+

GND

(Optional)

Description

HFORCE

Carries the signal current source.
Connected to the + side of the device
under test.

HSENSE

Together with Lsense, monitors the
Potential. Connected to the + side of
the device under test.

LSENSE

Together with Hsense, monitors the
Potential. Connected to the – side of
the device under test.

LFORCE

Accepts the signal current return.
Connected to the – side of the device
under test.

GND

If the test component has a large
metal area NOT connected to either
of the terminals, connect the GND
clip to minimize noise level.