Fixture connection – GW Instek LCR-8000G Series User Manual
Page 12
GETTING STARTED
23
23
23
23
Select AC mains frequency (50/60Hz)
Background
Although the LCR-8000G works under both 50 and
60Hz power frequencies, we recommend selecting
the frequency that matches the local setting to get
the best measurement precision, especially at
lower frequencies (< 100Hz).
Panel operation
1. Press the Menu key, then F5
(System). The system menu appears.
Menu
F 5
2. Press the Up/Down key and move
the cursor to Line frequency.
:
Line frequency
50Hz
3. If necessary, press the Left/Right
key to select the frequency, from 50
to 60Hz or from 60Hz to 50Hz.
:
Line frequency
60Hz
LCR-8000G Series User Manual
24
24
24
24
Fixture Connection
Fixture structure
Background
The standard fixture is a four-wire type with a
common terminal for screen connection. The outer
terminals (Hforce and Lforce) provide the current
and the inner terminals (Hsense and Lsense)
measures the potential.
Diagram
L
FORCE
L
SENSE
H
SENSE
H
FORCE
+
–
GND
(Optional)
Description
HFORCE
Carries the signal current source.
Connected to the + side of the device
under test.
HSENSE
Together with Lsense, monitors the
Potential. Connected to the + side of
the device under test.
LSENSE
Together with Hsense, monitors the
Potential. Connected to the – side of
the device under test.
LFORCE
Accepts the signal current return.
Connected to the – side of the device
under test.
GND
If the test component has a large
metal area NOT connected to either
of the terminals, connect the GND
clip to minimize noise level.