GW Instek LCR-8000G Series User Manual
Page 50

GRAPH MODE
99
99
99
99
Set vertical axis (Auto + Percentage)
Background
The Y (vertical) axis configuration is available for:
•
Manual or Automatic fit: Selects whether the
vertical range is manually set or automatically
adjusted as the graph is plotted.
•
Absolute or Percentage: Selects how the vertical
range is defined, as absolute values (minimum
and maximum) or percentage of the nominal
(center) value.
Panel operation
1. Press F2 to select % (Percentage).
F 2
2. Press F3 to select Auto Fit.
Auto Fit
F 3
3. The nominal value appears on the display.
Step Size:
2 4 8
1
Nominal:
1.0000mF
Set Nominal level 4. Press the Up/Down key to move the
cursor to Nominal.
Nominal:
1.0000mF
5. Use the numerical keys to enter the nominal
value on which the Hi and Lo percentages are
based.
Range
According to each measurement item
(see page37).
1.2mΩ
m
Ω
Enter
1.5kH
Enter
H
k
LCR-8000G Series User Manual
100
100
100
100
Backspace
All clear
Clear
If a wrong unit is entered, the value is cancelled.
Unit Mismatched
If a value outside of the range is entered, the
closest available value is automatically selected.
Nearest Available
6. The LCR-8000G automatically configures the
percentage (below and above the nominal level)
of the vertical scale.