BASIC MEASUREMENT

43

43

43

43

Reactance (X) and Susceptance (B = 1/X)

Background

Reactance measures the imaginary part of
Impedance (Z) caused by capacitors or inductors.
Susceptance is the reciprocal of Reactance and
measures the imaginary part of Admittance (Y),
which is also the reciprocal of Impedance.

Type

Series Reactance (X

S

)

Note: Reactance is
available only in series
circuit model.

Parallel Susceptance (B

P

)

Note: Susceptance is
available only in parallel
circuit model.

Display Range

0.01mΩ ~ 1GΩ

0.001ns ~ 1ks

Measurement
combination

X

S

+ Q

X

S

+ D

X

S

+ R

S

B

P

+ Q

B

P

+ D

B

P

+ R

P

B

P

+ G

P

Formula

θ

sin

1

Z

B

X

=

=

(

)

2

2

X

R

Z

S

+

=

(

)

2

2

X

R

RX

Z

P

+

=

θ

sin

Z

X

S

=

θ

sin

1

Y

X

B

=

=

(

)

2

2

B

G

GB

Y

S

+

=

(

)

2

2

B

G

Y

P

+

=

θ

sin

Y

B

P

=

LCR-8000G Series User Manual

44

44

44

44

Impedance (Z) and Admittance (Y = 1/Z)

Background

Impedance measures the total amount of
opposition between two terminals in an AC circuit.
Admittance is the reciprocal of Impedance and
measures how easily the electricity flows in an AC
circuit.

Type

Impedance (Z)

Admittance (Y)

Display Range

0.01mΩ ~ 1GΩ

0.001ns ~ 1ks

Formula

Y

I

E

Z

1

=

=

jX

R

Z

S

+

=

C

j

R

L

j

R

ϖ

ϖ

=

+

=

(

)

2

2

X

R

Z

S

+

=

(

)

2

2

X

R

RX

Z

P

+

=

θ

cos

Z

R

S

=

θ

sin

Z

X

S

=

Z

E

I

Y

1

=

=

jB

G

Y

P

+

=

L

j

G

C

j

G

ϖ

ϖ

=

+

=

(

)

2

2

B

G

GB

Y

S

+

=

(

)

2

2

B

G

Y

P

+

=

θ

cos

Y

G

P

=

θ

sin

Y

B

P

=