GW Instek LCR-8000G Series User Manual
Page 31
PASS-FAIL MODE
61
61
61
61
R
P
Parallel resistance θ
Angle
For detailed description of each item, see page37.
Configure beep setting
Background
The beep sounds when the pass/fail test result
matches the setting: failed or passed.
Panel operation
1. Press the Menu key, then F5
(System). The system configuration
appears.
Menu
F 5
2. Press the Up/Down key and move
the cursor to Beep.
Beep
:
OFF
3. Press the Left/Right key to select the
beep setting, Off, Pass, or Fail.
Off
Beep is turned Off
Pass
Beeps when the test result is pass
Fail
Beeps when the test result is fail
LCR-8000G Series User Manual
62
62
62
62
Beep in repetitive
mode
If the repetitive measurement is On, the beep
might sound continuously. If this becomes a
problem, either use the Single mode (press
Sing/Rep key) or turn Off the beep.
Configure the Average
Background
The Average function sets the number of samples
used, which are then averaged as the final output.
The number of samples varies from 1 to 256.
Panel operation
1. Press the Menu key, then F5
(System). The system configuration
appears.
Menu
F 5
2. Press the Up/Down key and move
the cursor to Average.
Average
:
10
3. Use the number pad to enter the number of
samples to be averaged. A maximum of 256
samples can be selected for averaging.
Average: 10
Enter