35

35

35

35

Wire capacitance When measuring the wire capacitance, the fixture

clips that are marked with H

F

(High Force)/H

S

(High Sense)

should always be connected to the

point that is influenced the most by noise.

Wire inductance The wire inductance should be subtracted from

the measurement result.

5cm, 1mm diameter wire has 50nH inductance

5cm, 2mm diameter wire has 40nH inductance

Frequency factor
in inductor
measurement

When an inductor is measured at a frequency
much lower than that for which it is designed (for
example, an HF choke tested at AF), the inductor
tends to behave as an inductive resistor. In these
circumstances, measurement accuracy is widened
by (1 + 1/Q) where Q is the quality factor.

Air-cored coils

Air-cored coils can pick up noise very easily,
therefore they should be kept well clear of any test
equipment that may contain power transformers
or display scan circuitry. Also, keep the coils away
from metal objects which may modify inductor
characteristics.

Iron-cored and
ferrite inductor

The effective value of iron-cored and ferrite
inductors can vary widely with the magnetization
and test signal level. Measure them at the AC level
and frequency in use. When core materials are
damaged by excessive magnetization (for
example: tape heads and microphone
transformers), check that the test signal is
acceptable before connection.

LCR-8000G Series User Manual

36

36

36

36

B

ASIC MEASUREMENT

Basic measurement measures DUT in numerical
style. Advanced measurements are available in
Pass/Fail test mode (page58), where measurement
results are compared with user-defined limits, and
in Graph mode (page87), where measurement
data is displayed in graphical representation.

Measurement
item

Measurement combination........................................ 37

Series/Parallel circuit models..................................... 38

Resistance (R) and Conductance (G = 1/R) .............. 40

Capacitance (C) .......................................................... 41

Inductance (L)............................................................. 42

Reactance (X) and Susceptance (B = 1/X)................. 43

Impedance (Z) and Admittance (Y = 1/Z)................. 44

Quality factor (Q) and Dissipation factor (D) ........... 45

Angle (θ) ..................................................................... 46

Measurement
overview

Enter measurement mode.......................................... 47

Display overview ......................................................... 48

Show circuit model or scale (pass/fail) ..................... 49

Measurement
configuration

Select measurement item........................................... 50

Set measurement range to Auto ................................ 51

Set measurement frequency....................................... 52

Set measurement voltage........................................... 54

Run measurement

Select Single measurement ........................................ 55

Select Repetitive measurement.................................. 56

Hide drive Voltage/Current ........................................ 57