BASIC MEASUREMENT

45

45

45

45

Quality factor (Q) and Dissipation factor (D)

Background

Both Quality factor and its reciprocal, Dissipation
factor, are used for measuring the rate of energy
dissipation relative to the measurement frequency.

Low energy dissipation: high Q, low D

High energy dissipation: low Q, high D

Type

Quality factor (Q)

Dissipation factor (D)

Display Range

0.01 ~ 9999.9

0.00001 ~ 1000

Formula

S

S

S

S

R

C

R

L

Q

ϖ

ϖ

1

=

=

P

P

P

P

R

C

L

R

ϖ

ϖ

=

=

(

)

D

1

90

tan

1

=

=

°

θ

S

S

S

S

R

C

L

R

D

ϖ

ϖ

=

=

P

P

P

P

G

L

C

G

ϖ

ϖ

=

=

(

)

Q

1

90

tan

=

=

°

θ

LCR-8000G Series User Manual

46

46

46

46

Angle (θ)

Background

The Angle (θ) measures the phase on which
Impedance (Z), Admittance (Y), Quality factor (Q),
and Dissipation factor (D) are measured.

Type

Angle (θ)

Display Range

–180° ~ +180°

Formula

jX

R

Z

S

+

=

C

j

R

L

j

R

ϖ

ϖ

=

+

=

(

)

D

Q

1

90

tan

1

=

=

°

θ

θ

cos

Z

R

S

=

θ

sin

Z

X

S

=

jB

G

Y

P

+

=

L

j

G

C

j

G

ϖ

ϖ

=

+

=

(

)

Q

D

1

90

tan

=

=

°

θ

θ

cos

Y

G

P

=

θ

sin

Y

B

P

=