Yokogawa PC-Based MX100 User Manual
Page 38
2-24
IM MX180-01E
On the channels of the input module, select from the modes below according to the input
signal. For details of the measurable range, measurement accuracy, and resolution, see
chapter 4, “Specification” in the MX100 Data Acquisition Unit User’s Manual (IM MX100-
01E).
• VOLT (DC voltage)
Select the measurement range from 20 mV, 60 mV, 200 mV, 2 V (default), 6 V, 20 V,
or 100 V.
Note
Current input is possible by attaching a shunt resistor to the input terminal and converting to voltage
input. The shunt resistors in the table below are available for purchase separately. For example, a
250 Ω shunt resistor is used to convert the signal in the range of 1 to 5 V for 4 to 20 mA input.
Part Name
Model
Resstance
Shunt resistor
(for clamp terminal)
415920
415921
415922
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
Shunt resistor
(for screw terminal)
438920
438921
438922
250 Ω ± 0.1%
100 Ω ± 0.1%
10 Ω ± 0.1%
• TC (thermocouple), on the Unversal Input Module and 30-CH, Medum-Speed
DCV/TC/DI Input Module
Select the thermocouple type (referred to as Range in the setup) from Type-R (default),
Type-S, Type-B, Type-K, Type-E, Type-J, Type-T, Type-N, Type-W, Type-L, Type-U, or
KpvsAu7Fe.
• RTD1 (resstance temperature detector, measured current: 1 mA), on the 4-CH,
Hgh-Speed Unversal Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high resolution Pt100), HQ JPt100 (high-resolution JPt100), Ni100:SAMA,
Ni100:DIN, or Ni120.
• RTD1 (resstance temperature detector, measured current of 1 mA), on the
10-CH, Medum-Speed Unversal Input Module/6-CH, Medum-Speed Four-Wre
RTD Resstance Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high-resolution Pt100), HQ JPt100 (high-resolution JPt100), Ni100:SAMA,
Ni100:DIN, Ni120, Pt50, Cu10:GE, Cu10:L&N, Cu10:WEED, Cu10:BAILEY, or J263B.
• RTD2 (resstance temperature detector, measured current: 2 mA), on the 4-CH,
Hgh-Speed Unversal Input Module
Select the RTD type (referred to as Range in the setup) from Pt100 (default), JPt100,
HQ Pt100 (high-resolution Pt100), HQ JPt100 (high-resolution JPt100), Pt50, Cu10:
GE, Cu10:L&N, Cu10:WEED, Cu10:BAILEY, or J263B.
The RTD2 selection is not available on the 10-CH, Medium-Speed Universal Input
Module and the 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module.
• RTDEX (resstance temperature detector, measured current: 0.25 mA), on the
6-CH, Medum-Speed Four-Wre RTD Resstance Input Module
Select the RTD type (referred to as Range in the setup) from Pt500 (default setting) or
Pt1000.
• OHM (resstance), on the 6-CH, Medum-Speed Four-Wre RTD Resstance Input
Module
Select a measurement range of 20 ohm (measured current: 1 mA), 200 ohm (measured
current: 1 mA), or 2 kohm (measured current: 0.25 mA).
2.4 Settng the Measurement Condtons (Settng the Measurement Channels)