Troubleshoot

ng

3

2

1

4

5

App

Index

The connected nput/output module s not detected.

Message

Correctve Acton

Reference

Page

Module connection or module startup error.

Attached the module while the power was ON.

Turn OFF the power. Detach the input/output module once and

attach it again.

*

The PC software, main module, and input/output

module style and release numbers do not

conform to the rule.

Check the release number of the PC software and the main

module style number, then upgrade the style.

[PC software release no.] ≥ [main module style no.] ≥

[input/output module style no.]

**

Carried out an incorrect calibration.

Recalibrate.

4-4

Connecting the 10-CH, Pulse Input Module

(MX114-PLS-M10).

The 10-CH Pulse Input Module cannot be used with the

MX100. Use this module with the MW100.

*

* See the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).
** See the MX100 Data Acquisition Unit Operation Guide (IM MX100-02E).

Alarms are not output.

Message

Correctve Acton

Reference

Page

There is a problem in the alarm setting.

Make the alarm and numerical output (relay) settings

appropriate.
Both the alarm and numerical output (relay) must be set.

2-54 to 2-59

Current s not output from the analog output module, and no output comes from the PWM output module.

Message

Correctve Acton

Reference

Page

External power supply required.

Connect an external power supply.

*

* See the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).

Transmsson output s not output from the analog output module, nor from the PWM output module.

Message

Correctve Acton

Reference

Page

“Transmission output execution” not enabled.

Enable it.

2-62

No output from the PWM output module

Message

Correctve Acton

Reference

Page

The output current limit (1 A) has been

exceeded. Once the limit is exceeded the

protection circuit activates and will not recover

until you turn OFF the external power.

Turn OFF the power. Check the load, then turn ON the power

again.

-

The fuse is blown. (4 A/module limit has been

exceeded)

Servicing required.

-

The CF card s not detected.

Message

Correctve Acton

Reference

Page

There is a problem with the CF card.

Replace the CF card.

Eject and format the CF card, then insert it again.

*

* See the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).

5.1 Troubleshootng