Explanaton, Calbrated parameters – Yokogawa PC-Based MX100 User Manual
Page 121
4-10
IM MX180-01E
4.2 Calbraton Procedure
Explanaton
Calbrated Parameters
• 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)
Calibration is carried out on a channel-by-channel basis.
Measurement Range to Be Calbrated
Input Value 1
Input Value 2
20 mV
0 mV
20 mV
60 mV
0 mV
60 mV
200 mV
0 mV
200 mV
1 V
0 V
1 V
2 V
0 V
2 V
6 V
0 V
6 V
20 V
0 V
20 V
100 V
0 V
100 V
RTD(1 mA)200 mV
0 Ω
200 Ω
RTD(1 mA)600 mV
0 Ω
300 Ω
RTD(1 mA)1 V
0 Ω
500 Ω
RTD(2 mA)60 mV
0 Ω
30 Ω
RTD(2 mA)200 mV
0 Ω
100 Ω
RTD(2 mA)600 mV
0 Ω
300 Ω
RTD(2 mA)1 V
0 Ω
250 Ω
• 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10)
Measurement Range to Be Calbrated
Input Value 1
Input Value 2
20 mV
0 mV on CH1
20 mV on CH2
60 mV
0 mV on CH1
60 mV on CH2
200 mV
0 mV on CH1
200 mV on CH2
1 V
0 V on CH1
1 V on CH2
2 V
0 V on CH1
2 V on CH2
6 V
0 V on CH1
6 V on CH2
20 V
0 V on CH1
20 V on CH2
100 V
0 V on CH1
100 V on CH2
RTD(1 mA)60 mV
0 Ω on CH3
60 Ω on CH6
RTD(1 mA)200 mV
0 Ω on CH3
200 Ω on CH4
RTD(1 mA)600 mV
0 Ω on CH3
300 Ω on CH5
• 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30)
Measurement Range to Be Calbrated
Input Value 1
Input Value 2
20 mV
0 mV on CH1
20 mV on CH2
60 mV
0 mV on CH1
60 mV on CH2
200 mV
0 mV on CH1
200 mV on CH2
1 V
0 V on CH1
1 V on CH2
2 V
0 V on CH1
2 V on CH2
6 V
0 V on CH1
6 V on CH2
20 V
0 V on CH1
20 V on CH2
100 V
0 V on CH1
100 V on CH2