4-10

IM MX180-01E

4.2 Calbraton Procedure

Explanaton

Calbrated Parameters

• 4-CH, High-Speed Universal Input Module (MX110-UNV-H04)
Calibration is carried out on a channel-by-channel basis.

Measurement Range to Be Calbrated

Input Value 1

Input Value 2

20 mV

0 mV

20 mV

60 mV

0 mV

60 mV

200 mV

0 mV

200 mV

1 V

0 V

1 V

2 V

0 V

2 V

6 V

0 V

6 V

20 V

0 V

20 V

100 V

0 V

100 V

RTD(1 mA)200 mV

0 Ω

200 Ω

RTD(1 mA)600 mV

0 Ω

300 Ω

RTD(1 mA)1 V

0 Ω

500 Ω

RTD(2 mA)60 mV

0 Ω

30 Ω

RTD(2 mA)200 mV

0 Ω

100 Ω

RTD(2 mA)600 mV

0 Ω

300 Ω

RTD(2 mA)1 V

0 Ω

250 Ω

• 10-CH, Medium-Speed Universal Input Module (MX110-UNV-M10)

Measurement Range to Be Calbrated

Input Value 1

Input Value 2

20 mV

0 mV on CH1

20 mV on CH2

60 mV

0 mV on CH1

60 mV on CH2

200 mV

0 mV on CH1

200 mV on CH2

1 V

0 V on CH1

1 V on CH2

2 V

0 V on CH1

2 V on CH2

6 V

0 V on CH1

6 V on CH2

20 V

0 V on CH1

20 V on CH2

100 V

0 V on CH1

100 V on CH2

RTD(1 mA)60 mV

0 Ω on CH3

60 Ω on CH6

RTD(1 mA)200 mV

0 Ω on CH3

200 Ω on CH4

RTD(1 mA)600 mV

0 Ω on CH3

300 Ω on CH5

• 30-CH, Medium-Speed DCV/TC/DI Input Module (MX110-VTD-L30)

Measurement Range to Be Calbrated

Input Value 1

Input Value 2

20 mV

0 mV on CH1

20 mV on CH2

60 mV

0 mV on CH1

60 mV on CH2

200 mV

0 mV on CH1

200 mV on CH2

1 V

0 V on CH1

1 V on CH2

2 V

0 V on CH1

2 V on CH2

6 V

0 V on CH1

6 V on CH2

20 V

0 V on CH1

20 V on CH2

100 V

0 V on CH1

100 V on CH2