4.2 Calbraton Procedure

• 6-CH, Medium-Speed Four-Wire RTD Resistance Input Module (MX110-V4R-M06)

Measurement Range to Be Calbrated

Input Value 1

Input Value 2

20 mV

0 mV on CH1

20 mV on CH2

60 mV

0 mV on CH1

60 mV on CH2

200 mV

0 mV on CH1

200 mV on CH2

1 V

0 V on CH1

1 V on CH2

2 V

0 V on CH1

2 V on CH2

6 V

0 V on CH1

6 V on CH2

20 V

0 V on CH1

20 V on CH2

100 V

0 V on CH1

100 V on CH2

RTD(1 mA)20 mV

0 Ω on CH3

20 Ω on CH4

RTD(1 mA)60 mV

0 Ω on CH3

60 Ω on CH5

RTD(1 mA)200 mV

0 Ω on CH3

200 Ω on CH6

RTD(1 mA)600 mV

0 Ω on CH3

300 Ω on CH4

RTD(0.25 mA)600 mV

0 Ω on CH3

2400 Ω on CH5

RTD(0.25 mA)1 V

0 Ω on CH3

3000 Ω on CH6

• 4-CH, Medium-Speed Strain Input Module (MX112-B12-M04, MX112-B35-M04,

MX112-NDI-M04)

Measurement Range to

Be Calbrated

Connecton 1

(Zero Calbraton)

Connecton 2

(Full Calbraton)

2000 µSTR (strain)

120.000 Ω on R1 to R4 of CH2 120.000 Ω on R1 to R3 of CH2

and 117.154 Ω on R4

20000 µSTR (strain)

120.000 Ω on R1 to R4 of CH2 120.000 Ω on R1 to R3 of CH2

and 113.010 Ω on R4

200000 µSTR (strain)

120.000 Ω on R1 to R4 of CH2 120.000 Ω on R1 to R3 of CH2

and 80,000 Ω on R4

Use a 4-gauge method connection when performing calibration. For information on

this connection, see section 2.4 of the MX100 Data Acquisition Unit User’s Manual (IM
MX100-01E).

• 8-CH, Medium-Speed Analog Output Module (MX120-VAO-M08)
Adjust so that all channels output 0 V and 10 V. This differs from other input modules.

Range to Be Calbrated

Output Value 1

Output Value 2

10 V

0 V

10 V

Note

• For a description of the measurement range and accuracy of each input module, see

chapter 4 in the MX100 Data Acquisition Unit User’s Manual (IM MX100-01E).

• For a description of the calibration procedure and wiring, see section 3.3 of the MX100 Data

Acquisition Unit User’s Manual (IM MX100-01E).