beautypg.com

2 calibration procedure, Calibration procedure -4, 2 calbraton procedure – Yokogawa PC-Based MX100 User Manual

Page 115: Ch10

background image

4-4

IM MX180-01E

4.2 Calbraton Procedure

Notes on Calbraton

• Do not perform other operations during calibration. Take extra care because if you

perform another operation while calibration is in progress, the module may suffer a
break down.

• When connecting the plate with screw terminals (model 772080) to the 10-CH,

Medium-Speed Universal Input module, note that the terminal arrangement is different
than when connecting clamped terminals.

With the calibration software, modules to which plates with screw terminals are

connected are recognized as clamp terminals, meaning that the wiring diagrams show
the wiring for the clamp terminals.

Therefore when wiring and calibrating, refer to the text describing the terminal function

and terminal symbols that show the type of signals input to each terminal that are
printed on the back of the terminal cover of the 10-CH, plate with screw terminals.

CH1

1
2

CH2

3
4

CH3

5
6

CH4

7
8

CH5

11 RTD

11 RTD

9

10

CH6

1
2

CH7

3
4

CH8

5
6

CH9

7
8

CH10

9

10

Note: The b terminals are common to all channels.

They are connected internally.

Procedure

1.

Click the illustration of the input/output module that you wish to calibrate.

Moving the pointer over

an module that can be

calibrated shows the word

“Calibrate”