Index – Yokogawa PC-Based MX100 User Manual
Page 97
4-9
IM MX100-01E
Specification
3
2
1
4
5
Index
The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).
Input
Measurement
Range Type
Rated Measurement
Range
Measurement Accuracy
Highest
Resolution
(1 Digit)
Integration Time: 16.67 ms
or More
Integration Time: 1.67 ms
DC voltage
60 mV (high res.)
0.000 to 60.000 mV
±(0.05% of rdg + 20 digits)
±(0.1% of rdg + 100 digits)
1 µV
1 V
–1.0000 to 1.0000 V
±(0.05% of rdg + 2 digits)
±(0.1% of rdg + 10 digits)
100 µV
6 V (high res.)
0.0000 to 6.0000 V
±(0.05% of rdg + 20 digits)
±(0.1% of rdg + 100 digits)
100 µV
Thermocouple
(excludes RJC accuracy
, when
burnout is OFF)
PLATINEL
0.0 to 1400.0°C
±(0.05% of rdg + 1°C)
±(0.1% of rdg + 4°C)
0.1°C
PR40-20
0.0 to 1900.0°C
±(0.05% of rdg + 2.5°C)
Except 300 to 700°C: ±6°C
less than 300°C: not guaranteed
±(0.1% of rdg + 12°C)
Except 300 to 700°C: ±25°C
less than 300°C: not guaranteed
NiNiMo
0.0 to 1310.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 2.7°C)
WRe3-25
0.0 to 2400.0°C
±(0.05% of rdg + 2°C)
Except 0 to 200°C: ±2.5°C
2000°C or higher:
±(0.05% of rdg + 4°C)
±(0.1% of rdg + 7°C)
Except 0 to 200°C: ±12°C
2000°C or higher:
±(0.1% of rdg + 11°C)
W/WRe26
0.0 to 2400.0°C
±(0.05% of rdg + 2°C)
Except 100 to 300°C: ±4°C
less than 100°C: not guaranteed
±(0.1% of rdg + 8.5°C)
Except 100 to 300°C: ±12°C
less than 100°C: not guaranteed
N (AWG14)
0.0 to 1300.0°C
±(0.05% of rdg + 0.7°C)
±(0.1% of rdg + 3.5°C)
XK GOST
–200.0 to 600.0°C
±(0.05% of rdg + 0.5°C)
Except –200 to 0°C:
±(0.2% of rdg + 0.7°C)
±(0.1% of rdg + 2.5°C)
Except –200 to 0°C:
±(1% of rdg + 2.5°C)
RTD
(Measurement
current: 1 mA)
Pt100
(noise resistance)
–200.0 to 600.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 2.5°C)
0.1°C
JPt100
(noise resistance)
–200.0 to 550.0°C
Pt100 GOST
–200.0 to 600.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
RTD
(Measurement
current: 2 mA)
Cu10 at 20°C
alpha = 0.00392
–200.0 to 300.0°C
±(0.1% of rdg + 0.7°C)
±(0.2% of rdg + 2.5°C)
Cu10 at 20°C
alpha = 0.00393
–200.0 to 300.0°C
Cu25 at 0°C
alpha = 0.00425
–200.0 to 300.0°C
±(0.1% of rdg + 0.5°C)
±(0.2% of rdg + 2°C)
Cu53 at 0°C
alpha = 0.00426035 –50.0 to 150.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
Cu100 at 0°C
alpha = 0.00425
–50.0 to 150.0°C
Pt25 (JPt100 × 1/4) –200.0 to 550.0°C
±(0.1% of rdg + 0.5°C)
±(0.2% of rdg + 2°C)
Cu10 GE
(high resolution)
–200.0 to 300.0°C
±(0.1% of rdg + 0.7°C)
±(0.2% of rdg + 2.5°C)
Cu10 L&N
(high resolution)
–200.0 to 300.0°C
Cu10 WEED
(high resolution)
–200.0 to 300.0°C
Cu10 BAILEY
(high resolution)
–200.0 to 300.0°C
Pt100
(noise resistance)
–200.0 to 250.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
JPt100
(noise resistance)
–200.0 to 550.0°C
Cu100 GOST
–200.0 to 200.0°C
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 1.5°C)
Cu50 GOST
–200.0 to 200.0°C
Cu10 GOST
–200.0 to 200.0°C
±(0.1% of rdg + 0.7°C)
±(0.2% of rdg + 2.5°C)
Measurement interval, integration time, and filter:
Measurement Interval Integration Time
Filter
Rejected Noise and Notes
10 ms
1.67 ms
Rectangular
600 Hz and its integer multiples*
50 ms
16.67 ms
60 Hz and its integer multiples
20 ms
50 Hz and its integer multiples
Auto
Automatically detects the power supply
frequency and set 16.67 or 20 ms
100, 200 ms
36.67 ms
Trapezoidal
50 Hz or 60 Hz and their integer multiples
500 ms
100 ms
Rectangular
10 Hz and its integer multiples
1, 2, 5,10, 20, 30, 60 s
200 ms
Cos
Fc = 5-Hz low-pass filter
*
Since the power supply frequency noise is not rejected, the measured vales may fluctuate especially
with temperature measurement. In such cases, set the measurement interval to 50 ms or higher.
Reference junction compensation:
Switch between external and internal on each channel, remote
RJC function available